Comparison of Strength, Consistency, and Speed of COR-KNOT versus Manually Hand-Tied Knots in an Ex Vivo Minimally Invasive Model. Issue 2 (March 2014)
- Record Type:
- Journal Article
- Title:
- Comparison of Strength, Consistency, and Speed of COR-KNOT versus Manually Hand-Tied Knots in an Ex Vivo Minimally Invasive Model. Issue 2 (March 2014)
- Main Title:
- Comparison of Strength, Consistency, and Speed of COR-KNOT versus Manually Hand-Tied Knots in an Ex Vivo Minimally Invasive Model
- Authors:
- Lee, Candice Y.
Sauer, Jude S.
Gorea, Heather R.
Martellaro, Angelo J.
Knight, Peter A. - Abstract:
- Objective: This study compared the strength, consistency, and speed of prosthetic attachment sutures secured with automated fasteners with those of manual knots using an ex vivo porcine mitral valve annuloplasty model. A novel miniature pressure transducer system was developed to quantify pressures between sutured prosthetic rings and underlying cardiac tissue. Methods: Sixteen mitral annuloplasty rings were sewn into ex vivo pig hearts. Eight rings were secured with the COR-KNOT device; and eight rings, with hand-tied knots using a knot pusher. A cardiac surgeon and a surgery resident each completed four manually tied rings and four COR-KNOT rings via a thoracotomy trainer. The total time to knot and cut each ring's sutures was recorded. Suture attachment pressures were measured within (intrasuture) and between (extrasuture) each suture loop using a 0.5 × 2.0-mm microtransducer probe system. Results: The suture holding pressures for the COR-KNOT fasteners were significantly greater than for the manually tied knots (median, 1008.9 vs 415.8 mm Hg, P < 0.001). All automated fasteners measured greater than 500 mm Hg, whereas 56% of the hand-tied knots were less than 500 mm Hg, and 14% were less than 75 mm Hg. There was less variation in attachment pressures for the COR-KNOT fasteners than for the hand-tied knots (SD, 401.6 vs 499.3 mm Hg, P = 0.04). Significant time savings occurred with the use of the COR-KNOT compared with manual tying (12.4 vs 71.1 seconds perknot, P =Objective: This study compared the strength, consistency, and speed of prosthetic attachment sutures secured with automated fasteners with those of manual knots using an ex vivo porcine mitral valve annuloplasty model. A novel miniature pressure transducer system was developed to quantify pressures between sutured prosthetic rings and underlying cardiac tissue. Methods: Sixteen mitral annuloplasty rings were sewn into ex vivo pig hearts. Eight rings were secured with the COR-KNOT device; and eight rings, with hand-tied knots using a knot pusher. A cardiac surgeon and a surgery resident each completed four manually tied rings and four COR-KNOT rings via a thoracotomy trainer. The total time to knot and cut each ring's sutures was recorded. Suture attachment pressures were measured within (intrasuture) and between (extrasuture) each suture loop using a 0.5 × 2.0-mm microtransducer probe system. Results: The suture holding pressures for the COR-KNOT fasteners were significantly greater than for the manually tied knots (median, 1008.9 vs 415.8 mm Hg, P < 0.001). All automated fasteners measured greater than 500 mm Hg, whereas 56% of the hand-tied knots were less than 500 mm Hg, and 14% were less than 75 mm Hg. There was less variation in attachment pressures for the COR-KNOT fasteners than for the hand-tied knots (SD, 401.6 vs 499.3 mm Hg, P = 0.04). Significant time savings occurred with the use of the COR-KNOT compared with manual tying (12.4 vs 71.1 seconds perknot, P = 0.001). Conclusions: The novel microtransducer technology provided an innovative means of evaluating cardiac prosthetic anchoring sutures. In this model, mitral annuloplasty ring sutures secured with the COR-KNOT device were stronger, more consistent, and faster than with manually tied knots. … (more)
- Is Part Of:
- Innovations. Volume 9:Issue 2(2014)
- Journal:
- Innovations
- Issue:
- Volume 9:Issue 2(2014)
- Issue Display:
- Volume 9, Issue 2 (2014)
- Year:
- 2014
- Volume:
- 9
- Issue:
- 2
- Issue Sort Value:
- 2014-0009-0002-0000
- Page Start:
- 111
- Page End:
- 116
- Publication Date:
- 2014-03
- Subjects:
- Minimally invasive surgery -- Valve surgery -- Hand-tied knots -- Automated titanium fasteners -- Pressure transducer
Cardiovascular system -- Surgery -- Periodicals
Heart -- Surgery -- Periodicals
Blood-vessels -- Surgery -- Periodicals
Chest -- Surgery -- Periodicals
Thoracic Surgical Procedures -- methods -- Periodicals
Vascular Surgical Procedures -- methods -- Periodicals
Thorax -- Chirurgie -- Méthodologie -- Périodiques
Vaisseaux sanguins -- Chirurgie -- Méthodologie -- Périodiques
Blood-vessels -- Surgery
Chest -- Surgery
Periodicals
617.41 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&NEWS=n&PAGE=toc&D=ovft&AN=01243895-000000000-00000 ↗
http://journals.lww.com/innovjournal/pages/default.aspx ↗
http://www.lww.com/product/?1556-9845 ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.1097/imi.0000000000000051 ↗
- Languages:
- English
- ISSNs:
- 1556-9845
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24249.xml