Development of a new HotBalloonTM ablation catheter equipped with a balloon surface temperature monitoring sensor for pulmonary vein isolation. (17th November 2022)
- Record Type:
- Journal Article
- Title:
- Development of a new HotBalloonTM ablation catheter equipped with a balloon surface temperature monitoring sensor for pulmonary vein isolation. (17th November 2022)
- Main Title:
- Development of a new HotBalloonTM ablation catheter equipped with a balloon surface temperature monitoring sensor for pulmonary vein isolation
- Authors:
- Tsukamoto, Kota
Tanahashi, Akio
Harada, Hiroyuki
Takaoka, Motoki
Sohara, Hiroshi - Abstract:
- Abstract: The balloon surface temperature (BST) should be monitored to ensure the success of the ablation procedure using the HotBalloon TM ablation catheter (HBC) in clinical settings. Therefore, we sought to develop a new HBC equipped with a surface temperature monitoring sensor. The BST was evaluated using a pseudo-tissue model and a thermocouple to imitate catheter insertion into the pulmonary vein. Thermo-fluid analysis with computer-aided engineering (CAE) was performed to analyse the temperature distribution in the catheter and the balloon. The CAE analysis reproduced the results from a pseudo-tissue model experiment and demonstrated that some fluid zones inside the catheter shaft had a nearly identical temperature as the BST during the liquid suction period. The pseudo-tissue model experiment confirmed that the temperature was almost the same between the balloon surface and the position of 5 mm inside the catheter shaft from the proximal end of the balloon. A thermocouple placed at 5 mm or 25 mm from the proximal end of the balloon within the catheter shaft showed an equivalent temperature result. This 5–25-mm distance is acceptable to set the BST monitoring sensor inside the catheter shaft, since the sensor can help accurately estimate the BST.
- Is Part Of:
- Journal of medical engineering & technology. Volume 46:Number 8(2022)
- Journal:
- Journal of medical engineering & technology
- Issue:
- Volume 46:Number 8(2022)
- Issue Display:
- Volume 46, Issue 8 (2022)
- Year:
- 2022
- Volume:
- 46
- Issue:
- 8
- Issue Sort Value:
- 2022-0046-0008-0000
- Page Start:
- 687
- Page End:
- 692
- Publication Date:
- 2022-11-17
- Subjects:
- Atrial fibrillation -- catheter ablation -- HotBalloon -- balloon surface temperature -- computer-aided engineering
Biomedical engineering -- Periodicals
Medical technology -- Periodicals
610.28 - Journal URLs:
- http://informahealthcare.com/journal/jmt ↗
http://www.tandfonline.com/toc/ijmt20/current ↗
http://informahealthcare.com ↗
http://www.tandf.co.uk/journals/titles/03091902.asp ↗ - DOI:
- 10.1080/03091902.2022.2094010 ↗
- Languages:
- English
- ISSNs:
- 0309-1902
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5017.057000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24243.xml