FLUIDIC PERFORMANCE OF A DUAL-ACTION VITRECTOMY PROBE COMPARED WITH A SINGLE-ACTION PROBE. Issue 11 (21st November 2022)
- Record Type:
- Journal Article
- Title:
- FLUIDIC PERFORMANCE OF A DUAL-ACTION VITRECTOMY PROBE COMPARED WITH A SINGLE-ACTION PROBE. Issue 11 (21st November 2022)
- Main Title:
- FLUIDIC PERFORMANCE OF A DUAL-ACTION VITRECTOMY PROBE COMPARED WITH A SINGLE-ACTION PROBE
- Authors:
- Steel, David H.
Charles, Martin
Zhu, Ying
Tambat, Sonalee
Irannejad, A. Mani
Charles, Steve - Abstract:
- Abstract : Supplemental Digital Content is Available in the Text. The dual-action 20, 000 cpm vitrectomy probes demonstrated a stable fluidic performance and greater aspiration flow, reduced nearfield effects, and lower traction, compared with the single-action 10, 000 cpm probes. Dual-action probes required less vacuum to achieve comparable flow rates. Abstract : Purpose: To assess flow rates, nearfield effects, and traction of a dual-cutting 20, 000 cpm vitrectomy probe (HYPERVIT, Alcon) versus a single-cutting 10, 000 cpm probe (Advanced ULTRAVIT, Alcon). Methods: Flow rates were evaluated for 25+ and 27+ gauge probes using balanced salt solution or porcine cadaver vitreous (biased open, 50/50, and biased closed duty cycles). Probes were suspended in an open beaker, and flow rates were calculated using a precision balance. Nearfield effects and flow pulsatility were assessed using a validated simulation model based on experimental microparticle image velocimetry. Traction was assessed by attaching vitreous to a cantilever beam and measuring the deflection of the beam. Results: For HYPERVIT probes, aqueous flow rates were similar across all cutting rates. Vitreous flow rates increased with increasing cutting rates. At maximum cutting rates, aqueous flow was 62%–67% greater (25+) and 63% greater (27+) with HYPERVIT versus Advanced ULTRAVIT ( P < 0.05); vitreous flow was 44%–47% greater (25+) and 26%–32% greater (27+) with HYPERVIT versus Advanced ULTRAVIT ( P < 0.05).Abstract : Supplemental Digital Content is Available in the Text. The dual-action 20, 000 cpm vitrectomy probes demonstrated a stable fluidic performance and greater aspiration flow, reduced nearfield effects, and lower traction, compared with the single-action 10, 000 cpm probes. Dual-action probes required less vacuum to achieve comparable flow rates. Abstract : Purpose: To assess flow rates, nearfield effects, and traction of a dual-cutting 20, 000 cpm vitrectomy probe (HYPERVIT, Alcon) versus a single-cutting 10, 000 cpm probe (Advanced ULTRAVIT, Alcon). Methods: Flow rates were evaluated for 25+ and 27+ gauge probes using balanced salt solution or porcine cadaver vitreous (biased open, 50/50, and biased closed duty cycles). Probes were suspended in an open beaker, and flow rates were calculated using a precision balance. Nearfield effects and flow pulsatility were assessed using a validated simulation model based on experimental microparticle image velocimetry. Traction was assessed by attaching vitreous to a cantilever beam and measuring the deflection of the beam. Results: For HYPERVIT probes, aqueous flow rates were similar across all cutting rates. Vitreous flow rates increased with increasing cutting rates. At maximum cutting rates, aqueous flow was 62%–67% greater (25+) and 63% greater (27+) with HYPERVIT versus Advanced ULTRAVIT ( P < 0.05); vitreous flow was 44%–47% greater (25+) and 26%–32% greater (27+) with HYPERVIT versus Advanced ULTRAVIT ( P < 0.05). Nearfield effects were reduced, and peak traction forces were significantly lower for HYPERVIT versus Advanced ULTRAVIT ( P < 0.05). Conclusion: Significantly greater aspiration flow, reduced nearfield effects, and reduced traction were observed with dual-action versus single-action probes. … (more)
- Is Part Of:
- Retina. Volume 42:Issue 11(2022)
- Journal:
- Retina
- Issue:
- Volume 42:Issue 11(2022)
- Issue Display:
- Volume 42, Issue 11 (2022)
- Year:
- 2022
- Volume:
- 42
- Issue:
- 11
- Issue Sort Value:
- 2022-0042-0011-0000
- Page Start:
- 2150
- Page End:
- 2158
- Publication Date:
- 2022-11-21
- Subjects:
- dual-action vitrectomy probes -- flow rates -- high-speed vitrectomy -- nearfield effects -- pulsatility -- traction
Retina -- Diseases -- Periodicals
Retinal Diseases
Vitreous Body
617.735 - Journal URLs:
- http://journals.lww.com/retinajournal/pages/default.aspx ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/IAE.0000000000003573 ↗
- Languages:
- English
- ISSNs:
- 0275-004X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7785.510300
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24196.xml