Water and lipid suppression techniques for advanced 1H MRS and MRSI of the human brain: Experts' consensus recommendations. (16th December 2020)
- Record Type:
- Journal Article
- Title:
- Water and lipid suppression techniques for advanced 1H MRS and MRSI of the human brain: Experts' consensus recommendations. (16th December 2020)
- Main Title:
- Water and lipid suppression techniques for advanced 1H MRS and MRSI of the human brain: Experts' consensus recommendations
- Authors:
- Tkáč, Ivan
Deelchand, Dinesh
Dreher, Wolfgang
Hetherington, Hoby
Kreis, Roland
Kumaragamage, Chathura
Považan, Michal
Spielman, Daniel M.
Strasser, Bernhard
de Graaf, Robin A. - Other Names:
- Choi In‐Young guestEditor.
Kreis Roland guestEditor. - Abstract:
- Abstract : The neurochemical information provided by proton magnetic resonance spectroscopy (MRS) or MR spectroscopic imaging (MRSI) can be severely compromised if strong signals originating from brain water and extracranial lipids are not properly suppressed. The authors of this paper present an overview of advanced water/lipid‐suppression techniques and describe their advantages and disadvantages. Moreover, they provide recommendations for choosing the most appropriate techniques for proper use. Methods of water signal handling are primarily focused on the VAPOR technique and on MRS without water suppression (metabolite cycling). The section on lipid‐suppression methods in MRSI is divided into three parts. First, lipid‐suppression techniques that can be implemented on most clinical MR scanners (volume preselection, outer‐volume suppression, selective lipid suppression) are described. Second, lipid‐suppression techniques utilizing the combination of k‐space filtering, high spatial resolutions and lipid regularization are presented. Finally, three promising new lipid‐suppression techniques, which require special hardware (a multi‐channel transmit system for dynamic B1 + shimming, a dedicated second‐order gradient system or an outer volume crusher coil) are introduced. Abstract : The neurochemical information provided by 1 H MRS or MRSI data can be severely compromised if strong signals originating from the brain water and extracranial lipids are not properly suppressed. TheAbstract : The neurochemical information provided by proton magnetic resonance spectroscopy (MRS) or MR spectroscopic imaging (MRSI) can be severely compromised if strong signals originating from brain water and extracranial lipids are not properly suppressed. The authors of this paper present an overview of advanced water/lipid‐suppression techniques and describe their advantages and disadvantages. Moreover, they provide recommendations for choosing the most appropriate techniques for proper use. Methods of water signal handling are primarily focused on the VAPOR technique and on MRS without water suppression (metabolite cycling). The section on lipid‐suppression methods in MRSI is divided into three parts. First, lipid‐suppression techniques that can be implemented on most clinical MR scanners (volume preselection, outer‐volume suppression, selective lipid suppression) are described. Second, lipid‐suppression techniques utilizing the combination of k‐space filtering, high spatial resolutions and lipid regularization are presented. Finally, three promising new lipid‐suppression techniques, which require special hardware (a multi‐channel transmit system for dynamic B1 + shimming, a dedicated second‐order gradient system or an outer volume crusher coil) are introduced. Abstract : The neurochemical information provided by 1 H MRS or MRSI data can be severely compromised if strong signals originating from the brain water and extracranial lipids are not properly suppressed. The authors of this paper present an overview of advanced water/lipid‐suppression techniques and describe their advantages and disadvantages. In addition, recommendations for choosing the most appropriate technique and for their proper usage are provided. … (more)
- Is Part Of:
- NMR in biomedicine. Volume 34:Number 5(2021)
- Journal:
- NMR in biomedicine
- Issue:
- Volume 34:Number 5(2021)
- Issue Display:
- Volume 34, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 34
- Issue:
- 5
- Issue Sort Value:
- 2021-0034-0005-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2020-12-16
- Subjects:
- crusher coils -- dynamic B1 shimming -- ECLIPSE -- high spatial resolution -- inversion recovery -- L‐regularization -- metabolite cycling -- VAPOR
Nuclear magnetic resonance -- Periodicals
Magnetic Resonance Spectroscopy -- Periodicals
574 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/nbm.4459 ↗
- Languages:
- English
- ISSNs:
- 0952-3480
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6113.931000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24174.xml