Axial length measurement and detection rates using a swept-source optical coherence tomography–based biometer in the presence of a dense vitreous hemorrhage. Issue 3 (March 2020)
- Record Type:
- Journal Article
- Title:
- Axial length measurement and detection rates using a swept-source optical coherence tomography–based biometer in the presence of a dense vitreous hemorrhage. Issue 3 (March 2020)
- Main Title:
- Axial length measurement and detection rates using a swept-source optical coherence tomography–based biometer in the presence of a dense vitreous hemorrhage
- Authors:
- Wang, Qinmei
Huang, Yingying
Gao, Rongrong
Chen, Hao
Song, Benhao
Tu, Ruixue
Zhao, Yune
McAlinden, Colm
Huang, Jinhai - Abstract:
- Abstract : The OA-2000 had a higher axial length measurement detection rate than IOLMaster and LENSTAR biometers in severe vitreous hemorrhage; it had a good agreement with IOLMaster, LENSTAR, and ultrasound measurements. Abstract : Purpose: To evaluate axial length (AL) measurements and detection rates using a swept-source optical coherence tomography (SS-OCT) biometer (OA-2000) compared with the IOLMaster, LENSTAR, and ultrasound (US) measurements in eyes with vitreous hemorrhage. Setting: Eye Hospital of Wenzhou Medical University, China. Design: Comparative evaluation of a diagnostic test or technology. Methods: Patients were adults with vitreous hemorrhage in one or both eyes. Lens Opacities Classification System III was used to determine they type and severity of cataract and the Forrester system was used to classify the grade of vitreous hemorrhage. The AL measurement was taken with the OA-2000, IOLMaster (v5.4), and LENSTAR (v2.1), and with US (Axis-II). The Bland-Altman test was used to assess the agreement between devices. Results: Forty eyes were scanned with each of the 4 biometers. For Forrester grade III, the detection rate was 100%, 41.7%, and 41.7% with the OA-2000, IOLMaster, and LENSTAR, respectively. For Forrester grades I and II, the detection rate was 46.4%, 3.57%, and 3.57% with the OA-2000, IOLMaster, and LENSTAR, respectively. The detection rate with US was 100% for all grades. The Bland-Altman limits of agreement between the 4 biometers indicated anAbstract : The OA-2000 had a higher axial length measurement detection rate than IOLMaster and LENSTAR biometers in severe vitreous hemorrhage; it had a good agreement with IOLMaster, LENSTAR, and ultrasound measurements. Abstract : Purpose: To evaluate axial length (AL) measurements and detection rates using a swept-source optical coherence tomography (SS-OCT) biometer (OA-2000) compared with the IOLMaster, LENSTAR, and ultrasound (US) measurements in eyes with vitreous hemorrhage. Setting: Eye Hospital of Wenzhou Medical University, China. Design: Comparative evaluation of a diagnostic test or technology. Methods: Patients were adults with vitreous hemorrhage in one or both eyes. Lens Opacities Classification System III was used to determine they type and severity of cataract and the Forrester system was used to classify the grade of vitreous hemorrhage. The AL measurement was taken with the OA-2000, IOLMaster (v5.4), and LENSTAR (v2.1), and with US (Axis-II). The Bland-Altman test was used to assess the agreement between devices. Results: Forty eyes were scanned with each of the 4 biometers. For Forrester grade III, the detection rate was 100%, 41.7%, and 41.7% with the OA-2000, IOLMaster, and LENSTAR, respectively. For Forrester grades I and II, the detection rate was 46.4%, 3.57%, and 3.57% with the OA-2000, IOLMaster, and LENSTAR, respectively. The detection rate with US was 100% for all grades. The Bland-Altman limits of agreement between the 4 biometers indicated an excellent agreement: OA-2000 vs IOLMaster (95% limits of agreement −0.15 to 0.05 mm), OA-2000 vs LENSTAR (−0.04 to 0.12 mm), and OA-2000 vs US (−0.3 to 0.56 mm). Conclusions: In vitreous hemorrhage, the detection rate with the SS-OCT OA-2000 biometer was better than that with the IOLMaster and LENSTAR. In cases in which all biometers were able to acquire an AL measurement an excellent agreement was found. … (more)
- Is Part Of:
- Journal of cataract and refractive surgery. Volume 46:Issue 3(2020)
- Journal:
- Journal of cataract and refractive surgery
- Issue:
- Volume 46:Issue 3(2020)
- Issue Display:
- Volume 46, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 46
- Issue:
- 3
- Issue Sort Value:
- 2020-0046-0003-0000
- Page Start:
- 360
- Page End:
- 364
- Publication Date:
- 2020-03
- Subjects:
- 617.7
- Journal URLs:
- http://journals.lww.com/pages/default.aspx ↗
- DOI:
- 10.1097/j.jcrs.0000000000000072 ↗
- Languages:
- English
- ISSNs:
- 0886-3350
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.900000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 24127.xml