Generation of 18F-FDG PET standard scan images from short scans using cycle-consistent generative adversarial network. (7th November 2022)
- Record Type:
- Journal Article
- Title:
- Generation of 18F-FDG PET standard scan images from short scans using cycle-consistent generative adversarial network. (7th November 2022)
- Main Title:
- Generation of 18F-FDG PET standard scan images from short scans using cycle-consistent generative adversarial network
- Authors:
- Ghafari, Ali
Sheikhzadeh, Peyman
Seyyedi, Negisa
Abbasi, Mehrshad
Farzenefar, Saeed
Yousefirizi, Fereshteh
Ay, Mohammad Reza
Rahmim, Arman - Abstract:
- Abstract: Objective. To improve positron emission tomography (PET) image quality, we aim to generate images of quality comparable to standard scan duration images using short scan duration (1/8 and 1/16 standard scan duration) inputs and assess the generated standard scan duration images quantitative and qualitatively. Also, the effect of training dataset properties (i.e. body mass index (BMI)) on the performance of the model(s) will be explored. Approach. Whole-body PET scans of 42 patients (41 18 F-FDG and one 68 Ga-PSMA) scanned with standard radiotracer dosage were included in this study. One 18 F-FDG patient data was set aside and the remaining 40 patients were split into four subsets of 10 patients with different mean patient BMI. Multiple copies of a developed cycle-GAN network were trained on each subset to predict standard scan images using 1/8 and 1/16 short duration scans. Also, the models' performance was tested on a patient scanned with the 68 Ga-PSMA radiotracer. Quantitative performance was tested using peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and normalized root mean squared error (NRMSE) metrics, and two nuclear medicine specialists analyzed images qualitatively. Main results. The developed cycle-GAN model improved the PSNR, SSIM, and NRMSE of the 1/8 and 1/16 short scan duration inputs both 18 F-FDG and 68 Ga-PSMA radiotracers. Although, quantitatively PSNR, SSIM, and NRMSE of the 1/16 scan duration level were improved moreAbstract: Objective. To improve positron emission tomography (PET) image quality, we aim to generate images of quality comparable to standard scan duration images using short scan duration (1/8 and 1/16 standard scan duration) inputs and assess the generated standard scan duration images quantitative and qualitatively. Also, the effect of training dataset properties (i.e. body mass index (BMI)) on the performance of the model(s) will be explored. Approach. Whole-body PET scans of 42 patients (41 18 F-FDG and one 68 Ga-PSMA) scanned with standard radiotracer dosage were included in this study. One 18 F-FDG patient data was set aside and the remaining 40 patients were split into four subsets of 10 patients with different mean patient BMI. Multiple copies of a developed cycle-GAN network were trained on each subset to predict standard scan images using 1/8 and 1/16 short duration scans. Also, the models' performance was tested on a patient scanned with the 68 Ga-PSMA radiotracer. Quantitative performance was tested using peak signal-to-noise ratio (PSNR), structural similarity index (SSIM), and normalized root mean squared error (NRMSE) metrics, and two nuclear medicine specialists analyzed images qualitatively. Main results. The developed cycle-GAN model improved the PSNR, SSIM, and NRMSE of the 1/8 and 1/16 short scan duration inputs both 18 F-FDG and 68 Ga-PSMA radiotracers. Although, quantitatively PSNR, SSIM, and NRMSE of the 1/16 scan duration level were improved more than 1/8 counterparts, however, the later were qualitatively more appealing. SUVmean and SUVmax of the generated images were also indicative of the improvements. The cycle-GAN model was much more capable in terms of image quality improvements and speed than the NLM denoising method. All results proved statistically significant using the paired-sample T-Test statistical test ( p -value < 0.05). Significance. Our suggested approach based on cycle-GAN could improve image quality of the 1/8 and 1/16 short scan-duration inputs through noise reduction both quantitively (PSNR, SSIM, NRMSE, SUVmean, and SUVmax) and qualitatively (contrast, noise, and diagnostic capability) to the level comparable to the standard scan-duration counterparts. The cycle-GAN model(s) had a similar performance on the 68 Ga-PSMA to the 18 F-FDG images and could improve the images qualitatively and quantitatively but requires more extensive study. Overall, images predicted from 1/8 short scan-duration inputs had the upper hand compared with 1/16 short scan-duration inputs. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 67:Number 21(2022)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 67:Number 21(2022)
- Issue Display:
- Volume 67, Issue 21 (2022)
- Year:
- 2022
- Volume:
- 67
- Issue:
- 21
- Issue Sort Value:
- 2022-0067-0021-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-11-07
- Subjects:
- positron emission tomography -- deep learning -- 18F-FDG -- denoising -- short-duration scan
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ac950a ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 24107.xml