Syringe Drivers: Incorrect Selection of Syringe Type from the Syringe Menu May Result in Significant Errors in Drug Delivery. Issue 4 (July 2014)
- Record Type:
- Journal Article
- Title:
- Syringe Drivers: Incorrect Selection of Syringe Type from the Syringe Menu May Result in Significant Errors in Drug Delivery. Issue 4 (July 2014)
- Main Title:
- Syringe Drivers: Incorrect Selection of Syringe Type from the Syringe Menu May Result in Significant Errors in Drug Delivery
- Authors:
- Tooke, L. J.
Howell, L. - Abstract:
- There have been many reported adverse incidents due to syringe driver use, most of which have been attributable to human error. In this paper we present a previously unreported, but potentially widespread practice which may result in significant over or under-delivery of medication. Even with the naked eye it is evident that syringes of equal volume have different dimensions and to quantify this we sectioned a range of syringes and measured the inner and outer dimensions. Extensive menus for syringe brand and volumes are available on syringe drivers, offering users greater flexibility. However, this feature also allows users to select an incorrect syringe brand with potential consequences for drug delivery. We measured outputs under all selectable permutations, to determine the degree of fluid delivery variation and discovered inaccuracies in volumes ranging from 10% under-delivery to 24% over-delivery. There is a wide variation in syringe metrics and complex syringe menus may increase errors, resulting in significant under or over-delivery of medication. Availability of more than one brand of syringe in a clinical area increases the risk of adverse drug delivery events. Systems need to be implemented to minimise the risk of adverse events.
- Is Part Of:
- Anaesthesia and intensive care. Volume 42:Issue 4(2014)
- Journal:
- Anaesthesia and intensive care
- Issue:
- Volume 42:Issue 4(2014)
- Issue Display:
- Volume 42, Issue 4 (2014)
- Year:
- 2014
- Volume:
- 42
- Issue:
- 4
- Issue Sort Value:
- 2014-0042-0004-0000
- Page Start:
- 467
- Page End:
- 472
- Publication Date:
- 2014-07
- Subjects:
- syringe driver -- medication errors -- adverse events -- infusion pump
Anesthesiology -- Periodicals
Intensive Care Units -- Periodicals
617.96 - Journal URLs:
- https://journals.sagepub.com/home/aic ↗
- DOI:
- 10.1177/0310057X1404200407 ↗
- Languages:
- English
- ISSNs:
- 0310-057X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 24011.xml