Individualised flow-controlled ventilation versus pressure-controlled ventilation in a porcine model of thoracic surgery requiring one-lung ventilation: A laboratory study. Issue 11 (21st November 2022)
- Record Type:
- Journal Article
- Title:
- Individualised flow-controlled ventilation versus pressure-controlled ventilation in a porcine model of thoracic surgery requiring one-lung ventilation: A laboratory study. Issue 11 (21st November 2022)
- Main Title:
- Individualised flow-controlled ventilation versus pressure-controlled ventilation in a porcine model of thoracic surgery requiring one-lung ventilation
- Authors:
- Spraider, Patrick
Martini, Judith
Abram, Julia
Putzer, Gabriel
Ranalter, Manuela
Mathis, Simon
Hell, Tobias
Barnes, Tom
Enk, Dietmar - Abstract:
- Abstract : BACKGROUND: Flow-controlled ventilation (FCV) enables precise determination of dynamic compliance due to a continuous flow coupled with direct tracheal pressure measurement. Thus, pressure settings can be adjusted accordingly in an individualised approach. OBJECTIVE: The aim of this study was to compare gas exchange of individualised FCV to pressure-controlled ventilation (PCV) in a porcine model of simulated thoracic surgery requiring one-lung ventilation (OLV). DESIGN: Controlled interventional trial conducted on 16 domestic pigs. SETTING: Animal operating facility at the Medical University of Innsbruck. INTERVENTIONS: Thoracic surgery was simulated with left-sided thoracotomy and subsequent collapse of the lung over a period of three hours. When using FCV, ventilation was performed with compliance-guided pressure settings. When using PCV, end-expiratory pressure was adapted to achieve best compliance with peak pressure adjusted to achieve a tidal volume of 6 ml kg −1 during OLV. MAIN OUTCOME MEASURES: Gas exchange was assessed by the Horowitz index (= P aO2 /FIO2 ) and CO2 removal by the P aCO2 value in relation to required respiratory minute volume. RESULTS: In the FCV group ( n = 8) normocapnia could be maintained throughout the OLV trial despite a significantly lower respiratory minute volume compared to the PCV group ( n = 8) (8.0 vs. 11.6, 95% confidence interval, CI -4.5 to -2.7 l min −1 ; P < 0.001), whereas permissive hypercapnia had to be acceptedAbstract : BACKGROUND: Flow-controlled ventilation (FCV) enables precise determination of dynamic compliance due to a continuous flow coupled with direct tracheal pressure measurement. Thus, pressure settings can be adjusted accordingly in an individualised approach. OBJECTIVE: The aim of this study was to compare gas exchange of individualised FCV to pressure-controlled ventilation (PCV) in a porcine model of simulated thoracic surgery requiring one-lung ventilation (OLV). DESIGN: Controlled interventional trial conducted on 16 domestic pigs. SETTING: Animal operating facility at the Medical University of Innsbruck. INTERVENTIONS: Thoracic surgery was simulated with left-sided thoracotomy and subsequent collapse of the lung over a period of three hours. When using FCV, ventilation was performed with compliance-guided pressure settings. When using PCV, end-expiratory pressure was adapted to achieve best compliance with peak pressure adjusted to achieve a tidal volume of 6 ml kg −1 during OLV. MAIN OUTCOME MEASURES: Gas exchange was assessed by the Horowitz index (= P aO2 /FIO2 ) and CO2 removal by the P aCO2 value in relation to required respiratory minute volume. RESULTS: In the FCV group ( n = 8) normocapnia could be maintained throughout the OLV trial despite a significantly lower respiratory minute volume compared to the PCV group ( n = 8) (8.0 vs. 11.6, 95% confidence interval, CI -4.5 to -2.7 l min −1 ; P < 0.001), whereas permissive hypercapnia had to be accepted in PCV ( P aCO2 5.68 vs. 6.89, 95% CI -1.7 to -0.7 kPa; P < 0.001). The Horowitz index was comparable in both groups but calculated mechanical power was significantly lower in FCV (7.5 vs. 22.0, 95% CI -17.2 to -11.8 J min −1 ; P < 0.001). CONCLUSIONS: In this porcine study FCV maintained normocapnia during OLV, whereas permissive hypercapnia had to be accepted in PCV despite a substantially higher minute volume. Reducing exposure of the lungs to mechanical power applied by the ventilator in FCV offers a possible advantage for this mode of ventilation in terms of lung protection. … (more)
- Is Part Of:
- European journal of anaesthesiology. Volume 39:Issue 11(2022)
- Journal:
- European journal of anaesthesiology
- Issue:
- Volume 39:Issue 11(2022)
- Issue Display:
- Volume 39, Issue 11 (2022)
- Year:
- 2022
- Volume:
- 39
- Issue:
- 11
- Issue Sort Value:
- 2022-0039-0011-0000
- Page Start:
- 885
- Page End:
- 894
- Publication Date:
- 2022-11-21
- Subjects:
- Anesthesiology -- Periodicals
Anesthesiology -- Periodicals
Anesthésiologie -- Périodiques
Anesthesiology
Periodicals
Electronic journals
617.96 - Journal URLs:
- http://journals.lww.com/ejanaesthesiology/pages/default.aspx ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1365-2346/issues ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=eja ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00003643-000000000-00000 ↗
http://journals.lww.com ↗
http://www.lww.com/Product/0265-0215 ↗ - DOI:
- 10.1097/EJA.0000000000001745 ↗
- Languages:
- English
- ISSNs:
- 0265-0215
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.722200
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 23987.xml