Practical guidelines of online MR-guided adaptive radiotherapy. (10th August 2022)
- Record Type:
- Journal Article
- Title:
- Practical guidelines of online MR-guided adaptive radiotherapy. (10th August 2022)
- Main Title:
- Practical guidelines of online MR-guided adaptive radiotherapy
- Authors:
- Okamoto, Hiroyuki
Igaki, Hiroshi
Chiba, Takahito
Shibuya, Keiko
Sakasai, Tatsuya
Jingu, Keiichi
Inaba, Koji
Kuroda, Kagayaki
Aoki, Shigeki
Tatsumi, Daisaku
Nakamura, Mitsuhiro
Kadoya, Noriyuki
Furuyama, Yoshinobu
Kumazaki, Yu
Tohyama, Naoki
Tsuneda, Masato
Nishioka, Shie
Itami, Jun
Onishi, Hiroshi
Shigematsu, Naoyuki
Uno, Takashi - Abstract:
- Abstract: The first magnetic resonance (MR)-guided radiotherapy system in Japan was installed in May 2017. Implementation of online MR-guided adaptive radiotherapy (MRgART) began in February 2018. Online MRgART offers greater treatment accuracy owing to the high soft-tissue contrast in MR-images (MRI), compared to that in X-ray imaging. The Japanese Society for Magnetic Resonance in Medicine (JSMRM), Japan Society of Medical Physics (JSMP), Japan Radiological Society (JRS), Japanese Society of Radiological Technology (JSRT), and Japanese Society for Radiation Oncology (JASTRO) jointly established the comprehensive practical guidelines for online MRgART. These guidelines propose the essential requirements for clinical implementation of online MRgART with respect to equipment, personnel, institutional environment, practice guidance, and quality assurance/quality control (QA/QC). The minimum requirements for related equipment and QA/QC tools, recommendations for safe operation of MRI system, and the implementation system are described. The accuracy of monitor chamber and detector in dose measurements should be confirmed because of the presence of magnetic field. The ionization chamber should be MR-compatible. Non-MR-compatible devices should be used in an area that is not affected by the static magnetic field (outside the five Gauss line), and their operation should be checked to ensure that they do not affect the MR image quality. Dose verification should be performed using anAbstract: The first magnetic resonance (MR)-guided radiotherapy system in Japan was installed in May 2017. Implementation of online MR-guided adaptive radiotherapy (MRgART) began in February 2018. Online MRgART offers greater treatment accuracy owing to the high soft-tissue contrast in MR-images (MRI), compared to that in X-ray imaging. The Japanese Society for Magnetic Resonance in Medicine (JSMRM), Japan Society of Medical Physics (JSMP), Japan Radiological Society (JRS), Japanese Society of Radiological Technology (JSRT), and Japanese Society for Radiation Oncology (JASTRO) jointly established the comprehensive practical guidelines for online MRgART. These guidelines propose the essential requirements for clinical implementation of online MRgART with respect to equipment, personnel, institutional environment, practice guidance, and quality assurance/quality control (QA/QC). The minimum requirements for related equipment and QA/QC tools, recommendations for safe operation of MRI system, and the implementation system are described. The accuracy of monitor chamber and detector in dose measurements should be confirmed because of the presence of magnetic field. The ionization chamber should be MR-compatible. Non-MR-compatible devices should be used in an area that is not affected by the static magnetic field (outside the five Gauss line), and their operation should be checked to ensure that they do not affect the MR image quality. Dose verification should be performed using an independent dose verification system that has been confirmed to be reliable through commissioning. This guideline proposes the checklists to ensure the safety of online MRgART. Successful clinical implementation of online MRgART requires close collaboration between physician, radiological technologist, nurse, and medical physicist. … (more)
- Is Part Of:
- Journal of radiation research. Volume 63:Number 5(2022)
- Journal:
- Journal of radiation research
- Issue:
- Volume 63:Number 5(2022)
- Issue Display:
- Volume 63, Issue 5 (2022)
- Year:
- 2022
- Volume:
- 63
- Issue:
- 5
- Issue Sort Value:
- 2022-0063-0005-0000
- Page Start:
- 730
- Page End:
- 740
- Publication Date:
- 2022-08-10
- Subjects:
- Magnetic resonance (MR)-guided -- adaptive radiotherapy (ART) -- quality assurance/quality control (QA/QC)
Radiology, Medical -- Periodicals
Radiobiology -- Periodicals
Radiation -- Periodicals
616.0757 - Journal URLs:
- http://bibpurl.oclc.org/web/15847 ↗
http://bibpurl.oclc.org/web/7828 ↗
http://www.journalarchive.jst.go.jp/english/jnltop_en.php?cdjournal=jrr1960 ↗
https://www.jstage.jst.go.jp/browse/jrr ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/jrr/rrac048 ↗
- Languages:
- English
- ISSNs:
- 0449-3060
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23938.xml