Development of high‐resolution YAP(Ce) x‐ray camera for the imaging of astatine‐211(At‐211) in small animals. Issue 11 (22nd September 2020)
- Record Type:
- Journal Article
- Title:
- Development of high‐resolution YAP(Ce) x‐ray camera for the imaging of astatine‐211(At‐211) in small animals. Issue 11 (22nd September 2020)
- Main Title:
- Development of high‐resolution YAP(Ce) x‐ray camera for the imaging of astatine‐211(At‐211) in small animals
- Authors:
- Nakanishi, Kouhei
Yamamoto, Seiichi
Watabe, Tadashi
Kaneda‐Nakashima, Kazuko
Shirakami, Yoshifumi
Ooe, Kazuhiro
Toyoshima, Atsushi
Shinohara, Atsushi
Teramoto, Takahiro
Hatazawa, Jun
Kamada, Kei
Yoshikawa, Akira - Abstract:
- Abstract : Purpose: Astatine‐211 (At‐211) is a promising alpha emitter for radionuclide therapy. High‐resolution in vivo imaging of At‐211 in small animals is needed for the development of At‐211 radiopharmaceuticals. For this purpose, we developed a low‐energy x‐ray camera using a thin YAlO3 :Ce (YAP(Ce)) plate to image the low‐energy x rays (73–87 keV) from the daughter radionuclide of At‐211 (Po‐211). Method: We optically coupled a 38 × 38 × 1‐mm YAP (Ce) plate to a position‐sensitive photomultiplier (PSPMT) to develop an imaging detector. A pinhole or a parallel hole collimator was attached to the imaging detector, and the performance was measured for 60‐keV gamma photons. With the developed x‐ray camera, we carried out imaging of a mouse that had been administered At‐211‐NaAt. Results: The intrinsic spatial resolution of the YAP (Ce) x‐ray camera was approximately 1.2 mm FWHM, and the energy resolution was 22% FWHM. With a 5‐mm‐thick parallel hole collimator, the spatial resolution was 3.8 mm FWHM with a sensitivity of 8 × 10 −4 at 10 mm, which is a typical distance from the surface of the collimator to a subject in mouse imaging. Using a 1‐mm diameter pinhole collimator, the spatial resolution was 1.8 mm FWHM with a sensitivity of 3.5 × 10 −4 at 10 mm from the collimator. In the mouse images measured by the developed x‐ray camera, we could clearly observe that the At‐211 accumulated in the thyroid gland and the stomach of the mouse. Conclusion: We concluded that theAbstract : Purpose: Astatine‐211 (At‐211) is a promising alpha emitter for radionuclide therapy. High‐resolution in vivo imaging of At‐211 in small animals is needed for the development of At‐211 radiopharmaceuticals. For this purpose, we developed a low‐energy x‐ray camera using a thin YAlO3 :Ce (YAP(Ce)) plate to image the low‐energy x rays (73–87 keV) from the daughter radionuclide of At‐211 (Po‐211). Method: We optically coupled a 38 × 38 × 1‐mm YAP (Ce) plate to a position‐sensitive photomultiplier (PSPMT) to develop an imaging detector. A pinhole or a parallel hole collimator was attached to the imaging detector, and the performance was measured for 60‐keV gamma photons. With the developed x‐ray camera, we carried out imaging of a mouse that had been administered At‐211‐NaAt. Results: The intrinsic spatial resolution of the YAP (Ce) x‐ray camera was approximately 1.2 mm FWHM, and the energy resolution was 22% FWHM. With a 5‐mm‐thick parallel hole collimator, the spatial resolution was 3.8 mm FWHM with a sensitivity of 8 × 10 −4 at 10 mm, which is a typical distance from the surface of the collimator to a subject in mouse imaging. Using a 1‐mm diameter pinhole collimator, the spatial resolution was 1.8 mm FWHM with a sensitivity of 3.5 × 10 −4 at 10 mm from the collimator. In the mouse images measured by the developed x‐ray camera, we could clearly observe that the At‐211 accumulated in the thyroid gland and the stomach of the mouse. Conclusion: We concluded that the YAP (Ce) x‐ray camera is useful for in vivo imaging of At‐211. … (more)
- Is Part Of:
- Medical physics. Volume 47:Issue 11(2020)
- Journal:
- Medical physics
- Issue:
- Volume 47:Issue 11(2020)
- Issue Display:
- Volume 47, Issue 11 (2020)
- Year:
- 2020
- Volume:
- 47
- Issue:
- 11
- Issue Sort Value:
- 2020-0047-0011-0000
- Page Start:
- 5739
- Page End:
- 5748
- Publication Date:
- 2020-09-22
- Subjects:
- 211At -- characteristic x rays -- imaging -- in vivo imaging -- YAP(Ce)
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.14455 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23828.xml