Flows and function of the Infant Flow neonatal continuous positive airway pressure device investigated with computational fluid dynamics. (19th August 2020)
- Record Type:
- Journal Article
- Title:
- Flows and function of the Infant Flow neonatal continuous positive airway pressure device investigated with computational fluid dynamics. (19th August 2020)
- Main Title:
- Flows and function of the Infant Flow neonatal continuous positive airway pressure device investigated with computational fluid dynamics
- Authors:
- Drevhammar, Thomas
Berg, Niclas
Nilsson, Kjell
Jonsson, Baldvin
Prahl Wittberg, Lisa - Abstract:
- Abstract: Aim: The first dedicated neonatal continuous positive airway pressure (CPAP) device using variable flow was the Infant Flow. The system was pressure stable with a low resistance to breathing. The aim of this study was to describe the flow and function of the Infant Flow geometry using simulated breathing and computational fluid dynamics. Method: The original Infant Flow geometry was used with simulated term infant breathing at three levels of CPAP. The large eddy simulation methodology was applied in combination with the WALE sub‐grid scale model. Results: The simulation fully resolved the flow phenomena in the Infant Flow geometry. The main flow feature during inspiration was support by gas entrainment and mixing. During expiration, the jet deflected towards the outlet with unstable impingement of the jet at the opposing edge. Conclusion: The proposed mechanism has previously been based on theoretical reasoning, and our results present the first detailed description of the Infant Flow. The pressure stability was based on a jet supporting inspiration by gas entrainment and then being deflected during exhalation. This confirmed previously assumed principles of function and flows within the geometry and provided a base for further developments.
- Is Part Of:
- Acta pædiatrica. Volume 110:Number 3(2021)
- Journal:
- Acta pædiatrica
- Issue:
- Volume 110:Number 3(2021)
- Issue Display:
- Volume 110, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 110
- Issue:
- 3
- Issue Sort Value:
- 2021-0110-0003-0000
- Page Start:
- 811
- Page End:
- 817
- Publication Date:
- 2020-08-19
- Subjects:
- continuous positive airway pressure -- computational fluid dynamics -- equipment design -- infant premature
Pediatrics -- Periodicals
Pediatrics
618.92 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1651-2227 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/apa.15502 ↗
- Languages:
- English
- ISSNs:
- 0803-5253
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0642.400000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 23776.xml