Late complications following permanent pacemaker implantation or elective unit replacement. Issue 3 (1st September 1998)
- Record Type:
- Journal Article
- Title:
- Late complications following permanent pacemaker implantation or elective unit replacement. Issue 3 (1st September 1998)
- Main Title:
- Late complications following permanent pacemaker implantation or elective unit replacement
- Authors:
- Harcombe, A A
Newell, S A
Ludman, P F
Wistow, T E
Sharples, L D
Schofield, P M
Stone, D L
Shapiro, L M
Cole, T
Petch, M C - Abstract:
- Abstract : Objective: To determine the rate of late complications following first implantation or elective unit replacement of a permanent pacemaker system. Design: Analysis of pacemaker data and complications prospectively acquired on a computerised database. Complications were studied over an 11 year period from January 1984 to December 1994. Setting: Tertiary referral cardiothoracic centre. Patients: Records of 2621 patients were analysed retrospectively. Main outcome measures: Complications requiring repeat procedures occurring more than six weeks after pacemaker implantation or elective unit replacement. Results: The overall rate of late complications was significantly lower after first implantation of a permanent pacemaker (34 cases, complication rate 1.4%, 95% confidence interval 0.9% to 1.9%) than after elective unit replacement (16 cases, complication rate 6.5% (3.3% to 9.7%). There were 20 cases of erosion, 18 infections, five electrode problems, and seven miscellaneous problems. Complications were more common with inexperienced operators (18.9% (6.0% to 31.8%)) than with experienced operators (0.9% (0.3% to 1.5%)). Conclusions: The incidence of late complications following pacemaker implantation is low and compares favourably with early complication rates. The majority are caused by erosion and infection. Patients who have undergone elective unit replacement are at particular risk.
- Is Part Of:
- Heart. Volume 80:Issue 3(1998)
- Journal:
- Heart
- Issue:
- Volume 80:Issue 3(1998)
- Issue Display:
- Volume 80, Issue 3 (1998)
- Year:
- 1998
- Volume:
- 80
- Issue:
- 3
- Issue Sort Value:
- 1998-0080-0003-0000
- Page Start:
- 240
- Page End:
- 244
- Publication Date:
- 1998-09-01
- Subjects:
- permanent pacemaker -- elective unit replacement -- late complications -- arrhythmias
Heart -- Diseases -- Treatment -- Periodicals
Cardiology -- Periodicals
616.12 - Journal URLs:
- http://www.bmj.com/archive ↗
http://heart.bmj.com ↗
http://www.heartjnl.com ↗ - DOI:
- 10.1136/hrt.80.3.240 ↗
- Languages:
- English
- ISSNs:
- 1355-6037
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23717.xml