Evaluation of a Blood Glucose Monitoring System with Automatic High- and Low-Pattern Recognition Software in Insulin-Using Patients: Pattern Detection and Patient-Reported Insights. (July 2013)
- Record Type:
- Journal Article
- Title:
- Evaluation of a Blood Glucose Monitoring System with Automatic High- and Low-Pattern Recognition Software in Insulin-Using Patients: Pattern Detection and Patient-Reported Insights. (July 2013)
- Main Title:
- Evaluation of a Blood Glucose Monitoring System with Automatic High- and Low-Pattern Recognition Software in Insulin-Using Patients: Pattern Detection and Patient-Reported Insights
- Authors:
- Grady, Mike
Campbell, Denise
MacLeod, Kirsty
Srinivasan, Aparna - Abstract:
- Background: This study aimed to evaluate the performance of a glucose pattern recognition tool incorporated in a blood glucose monitoring system (BGMS) and its association with clinical measures, and to assess user perception and understanding of the pattern messages they receive. Methods: Participants had type 1 or type 2 diabetes mellitus and were self-adjusting insulin doses for ≥1 year. During a 4-week home testing period, participants performed ≥6 daily self-tests, adjusted their insulin regimen based on BGMS results, and recorded pattern messages in the logbook. Participants reflected on usability of the pattern tool in a questionnaire. Results: Study participants ( n = 101) received a mean ± standard deviation of 4.5 ± 1.9 pattern messages per week (3.6 ± 1.8 high glucose patterns and 0.9 ± 1.3 low glucose patterns). Most received ≥1 high (96.5%) and/or ≥1 low (46.0%) pattern message per week. The average number of high- and low-pattern messages per week was associated with higher and lower, respectively, baseline hemoglobin A1c ( p < .01) and fasting plasma glucose ( p < .05). Participants found high- and low-pattern messages clear and easy to understand (84.2% and 83.2%, respectively) and considered the frequency of low (82.0%) and high (63.4%) pattern messages about right. Overall, 71.3% of participants indicated they preferred to use a meter with pattern messages. Conclusions: The on-device Pattern tool identified meaningful blood glucose patterns, highlightingBackground: This study aimed to evaluate the performance of a glucose pattern recognition tool incorporated in a blood glucose monitoring system (BGMS) and its association with clinical measures, and to assess user perception and understanding of the pattern messages they receive. Methods: Participants had type 1 or type 2 diabetes mellitus and were self-adjusting insulin doses for ≥1 year. During a 4-week home testing period, participants performed ≥6 daily self-tests, adjusted their insulin regimen based on BGMS results, and recorded pattern messages in the logbook. Participants reflected on usability of the pattern tool in a questionnaire. Results: Study participants ( n = 101) received a mean ± standard deviation of 4.5 ± 1.9 pattern messages per week (3.6 ± 1.8 high glucose patterns and 0.9 ± 1.3 low glucose patterns). Most received ≥1 high (96.5%) and/or ≥1 low (46.0%) pattern message per week. The average number of high- and low-pattern messages per week was associated with higher and lower, respectively, baseline hemoglobin A1c ( p < .01) and fasting plasma glucose ( p < .05). Participants found high- and low-pattern messages clear and easy to understand (84.2% and 83.2%, respectively) and considered the frequency of low (82.0%) and high (63.4%) pattern messages about right. Overall, 71.3% of participants indicated they preferred to use a meter with pattern messages. Conclusions: The on-device Pattern tool identified meaningful blood glucose patterns, highlighting potential opportunities for improving glycemic control in patients who self-adjust their insulin. … (more)
- Is Part Of:
- Journal of diabetes science and technology. Volume 7:Number 4(2013)
- Journal:
- Journal of diabetes science and technology
- Issue:
- Volume 7:Number 4(2013)
- Issue Display:
- Volume 7, Issue 4 (2013)
- Year:
- 2013
- Volume:
- 7
- Issue:
- 4
- Issue Sort Value:
- 2013-0007-0004-0000
- Page Start:
- 970
- Page End:
- 978
- Publication Date:
- 2013-07
- Subjects:
- blood glucose monitoring system -- diabetes -- pattern analysis -- self-monitoring of blood glucose
Diabetes -- Periodicals
Medical technology -- Periodicals
Diabetes Mellitus -- Periodicals
616.462005 - Journal URLs:
- http://ejournals.ebsco.com/direct.asp?JournalID=712321 ↗
http://www.jodsat.org/about.html ↗
http://online.sagepub.com/ ↗ - DOI:
- 10.1177/193229681300700419 ↗
- Languages:
- English
- ISSNs:
- 1932-2968
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23701.xml