Quantification of Osseointegration of Plasma-Polymer Coated Titanium Alloyed Implants by means of Microcomputed Tomography versus Histomorphometry. (30th April 2015)
- Record Type:
- Journal Article
- Title:
- Quantification of Osseointegration of Plasma-Polymer Coated Titanium Alloyed Implants by means of Microcomputed Tomography versus Histomorphometry. (30th April 2015)
- Main Title:
- Quantification of Osseointegration of Plasma-Polymer Coated Titanium Alloyed Implants by means of Microcomputed Tomography versus Histomorphometry
- Authors:
- Gabler, Carolin
Zietz, Carmen
Bieck, Richard
Göhler, Rebecca
Lindner, Tobias
Haenle, Maximilian
Finke, Birgit
Meichsner, Jürgen
Testrich, Holger
Nowottnick, Mathias
Frerich, Bernhard
Bader, Rainer - Other Names:
- Deligianni Despina Academic Editor.
- Abstract:
- Abstract : A common method to derive both qualitative and quantitative data to evaluate osseointegration of implants is histomorphometry. The present study describes a new image reconstruction algorithm comparing the results of bone-to-implant contact (BIC) evaluated by means of µ CT with histomorphometry data. Custom-made conical titanium alloyed (Ti6Al4V) implants were inserted in the distal tibial bone of female Sprague-Dawley rats. Different surface configurations were examined: Ti6Al4V implants with plasma-polymerized allylamine (PPAAm) coating and plasma-polymerized ethylenediamine (PPEDA) coating as well as implants without surface coating. After six weeks postoperatively, tibiae were explanted and BIC was determined by µ CT (3D) and afterwards by histomorphometry (2D). In comparison to uncoated Ti6Al4V implants demonstrating low BIC of 32.4% (histomorphometry) and 51.3% ( µ CT), PPAAm and PPEDA coated implants showed a nonsignificant increase in BIC (histomorphometry: 45.7% and 53.5% and µ CT: 51.8% and 62.0%, resp.). Mean BIC calculated by µ CT was higher for all surface configurations compared to BIC detected by histomorphometry. Overall, a high correlation coefficient of 0.70 (p < 0.002 ) was found between 3D and 2D quantification of BIC. The μ CT analysis seems to be suitable as a nondestructive and accurate 3D imaging method for the evaluation of the bone-implant interface.
- Is Part Of:
- BioMed research international. Volume 2015(2015)
- Journal:
- BioMed research international
- Issue:
- Volume 2015(2015)
- Issue Display:
- Volume 2015, Issue 2015 (2015)
- Year:
- 2015
- Volume:
- 2015
- Issue:
- 2015
- Issue Sort Value:
- 2015-2015-2015-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-04-30
- Subjects:
- Medicine -- Periodicals
Biology -- Periodicals
Biotechnology -- Periodicals
Life sciences -- Periodicals
610.5 - Journal URLs:
- https://www.hindawi.com/journals/bmri/ ↗
- DOI:
- 10.1155/2015/103137 ↗
- Languages:
- English
- ISSNs:
- 2314-6133
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 23498.xml