Eryaman, Yigitcan - Abstract:
- Abstract : Purpose: In this study, we investigate a strategy to reduce the local specific absorption rate (SAR) while keeping B 1 + constant inside the region of interest (ROI) at the ultra‐high field (B0 ≥ 7T) MRI. Methods: Locally raising the resonance structure under the discontinuity (i.e., creating a bump) increases the distance between the accumulated charges and the tissue. As a result, it reduces the electric field and local SAR generated by these charges inside the tissue. The B 1 + at a point that is sufficiently far from the coil, however, is not affected by this modification. In this study, three different resonant elements (i.e., loop coil, snake antenna, and fractionated dipole [FD]) are investigated. For experimental validation, a bumped FD is further investigated at 10.5T. After the validation, the transmit performances of eight‐channel arrays of each element are compared through electromagnetic (EM) simulations. Results: Introducing a bump reduced the peak 10g‐averaged SAR by 21, 26, 23% for the loop and snake antenna at 7T, and FD at 10.5T, respectively. In addition, eight‐channel bumped FD array at 10.5T had a 27% lower peak 10g‐averaged SAR in a realistic human body simulation (i.e., prostate imaging) compared to an eight‐channel FD array. Conclusion: In this study, we investigated a simple design strategy based on adding bumps to a resonant element to reduce the local SAR while maintaining B 1 + inside an ROI. As an example, we modified an FD andAbstract : Purpose: In this study, we investigate a strategy to reduce the local specific absorption rate (SAR) while keeping B 1 + constant inside the region of interest (ROI) at the ultra‐high field (B0 ≥ 7T) MRI. Methods: Locally raising the resonance structure under the discontinuity (i.e., creating a bump) increases the distance between the accumulated charges and the tissue. As a result, it reduces the electric field and local SAR generated by these charges inside the tissue. The B 1 + at a point that is sufficiently far from the coil, however, is not affected by this modification. In this study, three different resonant elements (i.e., loop coil, snake antenna, and fractionated dipole [FD]) are investigated. For experimental validation, a bumped FD is further investigated at 10.5T. After the validation, the transmit performances of eight‐channel arrays of each element are compared through electromagnetic (EM) simulations. Results: Introducing a bump reduced the peak 10g‐averaged SAR by 21, 26, 23% for the loop and snake antenna at 7T, and FD at 10.5T, respectively. In addition, eight‐channel bumped FD array at 10.5T had a 27% lower peak 10g‐averaged SAR in a realistic human body simulation (i.e., prostate imaging) compared to an eight‐channel FD array. Conclusion: In this study, we investigated a simple design strategy based on adding bumps to a resonant element to reduce the local SAR while maintaining B 1 + inside an ROI. As an example, we modified an FD and performed EM simulations and phantom experiments with a 10.5T scanner. Results show that the peak 10g‐averaged SAR can be reduced more than 25%. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 84:Issue 6(2020)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 84:Issue 6(2020)
- Issue Display:
- Volume 84, Issue 6 (2020)
- Year:
- 2020
- Volume:
- 84
- Issue:
- 6
- Issue Sort Value:
- 2020-0084-0006-0000
- Page Start:
- 3485
- Page End:
- 3493
- Publication Date:
- 2020-06-23
- Subjects:
- 10.5 Tesla -- bumped transmitter -- MRI -- radiofrequency safety -- ultra‐high field
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.28382 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23461.xml