Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast‐enhanced breast MRI: Chemical shift artifact of the third kind. Issue 4 (24th August 2017)
- Record Type:
- Journal Article
- Title:
- Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast‐enhanced breast MRI: Chemical shift artifact of the third kind. Issue 4 (24th August 2017)
- Main Title:
- Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast‐enhanced breast MRI: Chemical shift artifact of the third kind
- Authors:
- Derakhshan, Jamal J.
McDonald, Elizabeth S.
Siegelman, Evan S.
Schnall, Mitchell D.
Wehrli, Felix W. - Abstract:
- Abstract : Purpose: To characterize errors in enhancement in breast dynamic contrast‐enhanced (DCE) MRI studies as a function of echo time and determine the source of dark band artifacts in clinical subtraction images. Methods: Computer simulations, oil and water substitute (methylene chloride), as well as an American College of Radiology quality control phantom were tested. Routine clinical DCE breast MRI study was bracketed with (accelerated) in‐phase DCE acquisitions in five patients. Results: Simulation results demonstrated up to −160% suppression of the expected enhancement caused by differential enhancement of fat and water. Two‐dimensional gradient‐recalled echo and fat‐suppressed 3D GRE phantom imaging confirmed the simulation results and showed that fat suppression does not eliminate the artifact. In vivo in‐phase DCE images showed increased enhancement consistent with predictions and also confirmed increased spatial blurring on in‐phase 3D gradient‐recalled echo images. Combined multi‐dimensional partial Fourier and parallel imaging provided a time‐equivalent in‐phase DCE MRI acquisition. Conclusion: Errors in expected enhancement occur in DCE breast MRI subtraction images because of differential enhancement of fat and water and incomplete fat signal suppression. These errors can lead to artificial suppression of enhancement as well as dark band artifacts on subtraction images. These artifacts can be eliminated with a time‐equivalent in‐phase fat‐suppressed 3DAbstract : Purpose: To characterize errors in enhancement in breast dynamic contrast‐enhanced (DCE) MRI studies as a function of echo time and determine the source of dark band artifacts in clinical subtraction images. Methods: Computer simulations, oil and water substitute (methylene chloride), as well as an American College of Radiology quality control phantom were tested. Routine clinical DCE breast MRI study was bracketed with (accelerated) in‐phase DCE acquisitions in five patients. Results: Simulation results demonstrated up to −160% suppression of the expected enhancement caused by differential enhancement of fat and water. Two‐dimensional gradient‐recalled echo and fat‐suppressed 3D GRE phantom imaging confirmed the simulation results and showed that fat suppression does not eliminate the artifact. In vivo in‐phase DCE images showed increased enhancement consistent with predictions and also confirmed increased spatial blurring on in‐phase 3D gradient‐recalled echo images. Combined multi‐dimensional partial Fourier and parallel imaging provided a time‐equivalent in‐phase DCE MRI acquisition. Conclusion: Errors in expected enhancement occur in DCE breast MRI subtraction images because of differential enhancement of fat and water and incomplete fat signal suppression. These errors can lead to artificial suppression of enhancement as well as dark band artifacts on subtraction images. These artifacts can be eliminated with a time‐equivalent in‐phase fat‐suppressed 3D gradient‐recalled echo sequence. Understanding chemical shift artifact of the third kind, a unique artifact of artificial enhancement suppression in the presence of intravoxel fat and water signal, will aid DCE breast MRI image interpretation. In‐phase acquisitions (combined with simultaneous minimum echo time or opposed‐phase echoes) may facilitate qualitative, quantitative and longitudinal analysis of contrast enhancement. Magn Reson Med 79:2277–2289, 2018. © 2017 International Society for Magnetic Resonance in Medicine. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 79:Issue 4(2018)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 79:Issue 4(2018)
- Issue Display:
- Volume 79, Issue 4 (2018)
- Year:
- 2018
- Volume:
- 79
- Issue:
- 4
- Issue Sort Value:
- 2018-0079-0004-0000
- Page Start:
- 2277
- Page End:
- 2289
- Publication Date:
- 2017-08-24
- Subjects:
- dynamic contrast‐enhanced (DCE) MRI -- subtraction artifact -- chemical shift artifact -- paradoxical enhancement -- breast MRI -- enhancement errors
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.26879 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23280.xml