Predictive Value of Transimpedance Matrix Measurements to Detect Electrode Tip Foldover. Issue 9 (30th October 2022)
- Record Type:
- Journal Article
- Title:
- Predictive Value of Transimpedance Matrix Measurements to Detect Electrode Tip Foldover. Issue 9 (30th October 2022)
- Main Title:
- Predictive Value of Transimpedance Matrix Measurements to Detect Electrode Tip Foldover
- Authors:
- Kay-Rivest, Emily
McMenomey, Sean O.
Jethanamest, Daniel
Shapiro, William H.
Friedmann, David R.
Waltzman, Susan B.
Roland, J. Thomas - Abstract:
- Abstract : Objective: To evaluate the ability of the transimpedance matrix (TIM) measurement to detect cochlear implant electrode tip foldover by comparing results to a "gold standard, " the intraoperative plain film radiograph. Study Design: Retrospective case series. Setting: Tertiary referral hospital. Patients: One hundred three patients who underwent cochlear implantation between June 2020 and August 2021. Interventions: Intraoperative electrophysiologic monitoring (electrode impedances, neural response telemetry, and TIM measurement) and modified Stenver's view plain film radiographs. Main Outcome Measures: Identification of tip foldover on both TIM and plain films. Results: In total, 103 patients (117 ears) had both a TIM measurement and intraoperative X-ray available for review, including 68 adults and 35 children. One hundred patients (85%) received the Cochlear Slim Modiolar electrode. Tip foldovers were noted in three of 117 implants (2.5%). In all cases, TIM was able to detect the foldover, and the electrode arrays were reinserted with the patients still under anesthesia, with repeat X-ray demonstrating a normal configuration. Two other abnormal TIM patterns were identified. One was in a patient with an obstructed cochlea in whom only 10 electrodes could be inserted, the other was in a patient with a common cavity abnormality. One additional patient underwent electrode repositioning intraoperatively because of overinsertion. In this patient, the TIM appeared toAbstract : Objective: To evaluate the ability of the transimpedance matrix (TIM) measurement to detect cochlear implant electrode tip foldover by comparing results to a "gold standard, " the intraoperative plain film radiograph. Study Design: Retrospective case series. Setting: Tertiary referral hospital. Patients: One hundred three patients who underwent cochlear implantation between June 2020 and August 2021. Interventions: Intraoperative electrophysiologic monitoring (electrode impedances, neural response telemetry, and TIM measurement) and modified Stenver's view plain film radiographs. Main Outcome Measures: Identification of tip foldover on both TIM and plain films. Results: In total, 103 patients (117 ears) had both a TIM measurement and intraoperative X-ray available for review, including 68 adults and 35 children. One hundred patients (85%) received the Cochlear Slim Modiolar electrode. Tip foldovers were noted in three of 117 implants (2.5%). In all cases, TIM was able to detect the foldover, and the electrode arrays were reinserted with the patients still under anesthesia, with repeat X-ray demonstrating a normal configuration. Two other abnormal TIM patterns were identified. One was in a patient with an obstructed cochlea in whom only 10 electrodes could be inserted, the other was in a patient with a common cavity abnormality. One additional patient underwent electrode repositioning intraoperatively because of overinsertion. In this patient, the TIM appeared to be within normal limits, but the over-insertion was apparent on X-ray. Overall, the sensitivity and specificity of TIM measurements in detecting electrode tip foldover were both 100%. Conclusion: TIM measurements were able to accurately identify tip foldovers. More research is needed to define the adjunctive role of TIM as an intraoperative measure. … (more)
- Is Part Of:
- Otology & neurotology. Volume 43:Issue 9(2022)
- Journal:
- Otology & neurotology
- Issue:
- Volume 43:Issue 9(2022)
- Issue Display:
- Volume 43, Issue 9 (2022)
- Year:
- 2022
- Volume:
- 43
- Issue:
- 9
- Issue Sort Value:
- 2022-0043-0009-0000
- Page Start:
- 1027
- Page End:
- 1032
- Publication Date:
- 2022-10-30
- Subjects:
- Cochlear implants -- Intraoperative monitoring -- Tip foldovers -- Transimpedance matrix
Otology -- Periodicals
Ear -- Diseases -- Periodicals
Skull base -- Surgery -- Periodicals
617.8005 - Journal URLs:
- http://www.otology-neurotology.com ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/MAO.0000000000003667 ↗
- Languages:
- English
- ISSNs:
- 1531-7129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6313.528000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 23259.xml