Radiofrequency excitation–related 23Na MRI signal loss in skeletal muscle, cartilage, and skin. Issue 6 (11th November 2019)
- Record Type:
- Journal Article
- Title:
- Radiofrequency excitation–related 23Na MRI signal loss in skeletal muscle, cartilage, and skin. Issue 6 (11th November 2019)
- Main Title:
- Radiofrequency excitation–related 23Na MRI signal loss in skeletal muscle, cartilage, and skin
- Authors:
- Kordzadeh, Atefeh
Duchscherer, Jade
Beaulieu, Christian
Stobbe, Rob - Abstract:
- Abstract : Purpose: To assess the sodium MRI signal loss resulting from typically used RF excitation pulses in human skeletal muscle, patellar cartilage, and skin. Methods: A double flip‐angle experiment was performed 3 times on the knees of 5 healthy volunteers with prescribed ω 1 = γ B1 of 1.67 kHz, 0.333 kHz, and 0.167 kHz. This was done to search for ω 1 ‐dependent increased rates of sodium‐23 central resonance flipping known to result from residual quadrupole splitting ( ωQ ), as this flip‐angle effect is associated with signal loss. This study facilitated in vivo regression of Gaussian‐distributed residual quadrupole splitting SD ( ωQ (SD) ) as well as T2fast and T2slow . Signal loss predicted from simulation was then compared with images acquired using 90° RF pulse lengths of 0.5 ms, 0.25 ms, and 0.15 ms. Results: Sodium‐23 central resonance flipping was significantly greater than prescribed (44% cartilage, 23% skin, 9% muscle) using ω 1 = 0.167 kHz, but only 4% cartilage, 5% skin, and 2% muscle using ω 1 = 1.67 kHz. Regression yielded ωQ (SD) = 420 ± 50 Hz for cartilage but no significant ωQ (SD) for skin or muscle. This points to rapid biexponential relaxation as the cause of the flip‐angle effect for skin/muscle. The T2fast(60%) /T2slow(40%) values were 1.6 ± 0.8 ms/16.1 ± 2.5 ms for muscle, 2.7 ± 0.9 ms/18.4 ± 2.5 ms for cartilage, and 0.4 ± 0.1 ms/9.3 ± 1.7 ms for skin. Simulation predicted signal loss of 6% ± 3%, 3% ± 1%, and 2% ± 1% for muscle, 16% ± 3%, 6% ±Abstract : Purpose: To assess the sodium MRI signal loss resulting from typically used RF excitation pulses in human skeletal muscle, patellar cartilage, and skin. Methods: A double flip‐angle experiment was performed 3 times on the knees of 5 healthy volunteers with prescribed ω 1 = γ B1 of 1.67 kHz, 0.333 kHz, and 0.167 kHz. This was done to search for ω 1 ‐dependent increased rates of sodium‐23 central resonance flipping known to result from residual quadrupole splitting ( ωQ ), as this flip‐angle effect is associated with signal loss. This study facilitated in vivo regression of Gaussian‐distributed residual quadrupole splitting SD ( ωQ (SD) ) as well as T2fast and T2slow . Signal loss predicted from simulation was then compared with images acquired using 90° RF pulse lengths of 0.5 ms, 0.25 ms, and 0.15 ms. Results: Sodium‐23 central resonance flipping was significantly greater than prescribed (44% cartilage, 23% skin, 9% muscle) using ω 1 = 0.167 kHz, but only 4% cartilage, 5% skin, and 2% muscle using ω 1 = 1.67 kHz. Regression yielded ωQ (SD) = 420 ± 50 Hz for cartilage but no significant ωQ (SD) for skin or muscle. This points to rapid biexponential relaxation as the cause of the flip‐angle effect for skin/muscle. The T2fast(60%) /T2slow(40%) values were 1.6 ± 0.8 ms/16.1 ± 2.5 ms for muscle, 2.7 ± 0.9 ms/18.4 ± 2.5 ms for cartilage, and 0.4 ± 0.1 ms/9.3 ± 1.7 ms for skin. Simulation predicted signal loss of 6% ± 3%, 3% ± 1%, and 2% ± 1% for muscle, 16% ± 3%, 6% ± 1%, and 3% ± 1% for cartilage, and 26% ± 7%, 15% ± 4%, and 10% ± 3% for skin when using 90° RF pulse lengths of 0.5 ms, 0.25 ms, and 0.15 ms, matching experiment. Conclusion: High‐power (short) RF pulses are necessary to reduce excitation‐related signal loss, particularly for sodium‐23 imaging of cartilage and skin. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 83:Issue 6(2020)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 83:Issue 6(2020)
- Issue Display:
- Volume 83, Issue 6 (2020)
- Year:
- 2020
- Volume:
- 83
- Issue:
- 6
- Issue Sort Value:
- 2020-0083-0006-0000
- Page Start:
- 1992
- Page End:
- 2001
- Publication Date:
- 2019-11-11
- Subjects:
- cartilage -- muscle -- quadrupole -- relaxation -- skin -- sodium MRI
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.28054 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23200.xml