A pilot study of quantitative capillary refill time to identify high blood lactate levels in critically ill patients. Issue 6 (19th August 2014)
- Record Type:
- Journal Article
- Title:
- A pilot study of quantitative capillary refill time to identify high blood lactate levels in critically ill patients. Issue 6 (19th August 2014)
- Main Title:
- A pilot study of quantitative capillary refill time to identify high blood lactate levels in critically ill patients
- Authors:
- Morimura, Naoto
Takahashi, Kohei
Doi, Tomoki
Ohnuki, Takahiro
Sakamoto, Tetsuya
Uchida, Yasuyuki
Takahashi, Hiroki
Fujita, Takashi
Ikeda, Hiroto - Abstract:
- Abstract : Introduction: We developed a new device to quantify capillary refill time (CRT) by applying the pulse oximeter principle, and evaluated the correlation between quantitative CRT (Q-CRT) and hypoperfusion status, as represented by blood lactate levels, in critically ill patients. Methods: A pilot study was undertaken in the intensive care unit (ICU) in a tertiary emergency medical centre. While the pulse oxygen saturation sensor was placed on the finger of the patients, transmitted light intensity (TLI) was measured with a pulse oximeter (OLV-3100; Nihon Kohden, Tokyo, Japan) before and during compression of the finger. Q-CRT was defined as the interval from the release of compression to the time when TLI reached 90% of baseline. Results: Q-CRT was analysed in a total of 57 waveforms among 23 patients and statistically correlated with lactate levels (Spearman's rank correlation coefficient, 0.681; p<0.001). The cut-off value of Q-CRT for predicting a lactate level of ≥2.0 mmol/L was 6.81 s (area under the curve (AUC) (95% CI 1.000 (1.000 to 1.000), p<0.001), and the value for predicting a lactate level of ≥4.0 mmol/L was 7.27 s (AUC=0.989 (95% CI 0.954 to 1.000), p<0.001). Conclusions: Q-CRT correlated with blood lactate levels in this pilot study. The most useful threshold for Q-CRT was ∼6–8 s. Further study is needed to investigate the potential role of this modality as a non-invasive predictor of hypoperfusion in the emergency department, ICU and operating roomAbstract : Introduction: We developed a new device to quantify capillary refill time (CRT) by applying the pulse oximeter principle, and evaluated the correlation between quantitative CRT (Q-CRT) and hypoperfusion status, as represented by blood lactate levels, in critically ill patients. Methods: A pilot study was undertaken in the intensive care unit (ICU) in a tertiary emergency medical centre. While the pulse oxygen saturation sensor was placed on the finger of the patients, transmitted light intensity (TLI) was measured with a pulse oximeter (OLV-3100; Nihon Kohden, Tokyo, Japan) before and during compression of the finger. Q-CRT was defined as the interval from the release of compression to the time when TLI reached 90% of baseline. Results: Q-CRT was analysed in a total of 57 waveforms among 23 patients and statistically correlated with lactate levels (Spearman's rank correlation coefficient, 0.681; p<0.001). The cut-off value of Q-CRT for predicting a lactate level of ≥2.0 mmol/L was 6.81 s (area under the curve (AUC) (95% CI 1.000 (1.000 to 1.000), p<0.001), and the value for predicting a lactate level of ≥4.0 mmol/L was 7.27 s (AUC=0.989 (95% CI 0.954 to 1.000), p<0.001). Conclusions: Q-CRT correlated with blood lactate levels in this pilot study. The most useful threshold for Q-CRT was ∼6–8 s. Further study is needed to investigate the potential role of this modality as a non-invasive predictor of hypoperfusion in the emergency department, ICU and operating room settings. … (more)
- Is Part Of:
- Emergency medicine journal. Volume 32:Issue 6(2015)
- Journal:
- Emergency medicine journal
- Issue:
- Volume 32:Issue 6(2015)
- Issue Display:
- Volume 32, Issue 6 (2015)
- Year:
- 2015
- Volume:
- 32
- Issue:
- 6
- Issue Sort Value:
- 2015-0032-0006-0000
- Page Start:
- 444
- Page End:
- 448
- Publication Date:
- 2014-08-19
- Subjects:
- Assessment -- Intensive Care -- Research, Clinical -- Clinical Assessment -- Equipment Evaluation
Emergency medicine -- Periodicals
616.02505 - Journal URLs:
- http://www.bmj.com/archive ↗
https://emj.bmj.com/ ↗ - DOI:
- 10.1136/emermed-2013-203180 ↗
- Languages:
- English
- ISSNs:
- 1472-0205
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23155.xml