Application of a novel diamond detector for commissioning of FLASH radiotherapy electron beams. Issue 8 (16th June 2022)
- Record Type:
- Journal Article
- Title:
- Application of a novel diamond detector for commissioning of FLASH radiotherapy electron beams. Issue 8 (16th June 2022)
- Main Title:
- Application of a novel diamond detector for commissioning of FLASH radiotherapy electron beams
- Authors:
- Verona Rinati, Gianluca
Felici, Giuseppe
Galante, Federica
Gasparini, Alessia
Kranzer, Rafael
Mariani, Giulia
Pacitti, Matteo
Prestopino, Giuseppe
Schüller, Andreas
Vanreusel, Verdi
Verellen, Dirk
Verona, Claudio
Marinelli, Marco - Abstract:
- Abstract: Purpose: A diamond detector prototype was recently proposed by Marinelli et al. (Medical Physics 2022, https://doi.org/10.1002/mp.15473 ) for applications in ultrahigh‐dose‐per‐pulse (UH‐DPP) and ultrahigh‐dose‐rate (UH‐DR) beams, as used in FLASH radiotherapy (FLASH‐RT). In the present study, such so‐called flashDiamond (fD) was investigated from the dosimetric point of view, under pulsed electron beam irradiation. It was then used for the commissioning of an ElectronFlash linac (SIT S.p.A., Italy) both in conventional and UH‐DPP modalities. Methods: Detector calibration was performed in reference conditions, under 60 Co and electron beam irradiation. Its response linearity was investigated in UH‐DPP conditions. For this purpose, the DPP was varied in the 1.2–11.9 Gy range, by changing either the beam applicator or the pulse duration from 1 to 4 μs. Dosimetric validation of the fD detector prototype was then performed in conventional modality, by measuring percentage depth dose (PDD) curves, beam profiles, and output factors (OFs). All such measurements were carried out in a motorized water phantom. The obtained results were compared with the ones from commercially available dosimeters, namely, a microDiamond, an Advanced Markus ionization chamber, a silicon diode detector, and EBT‐XD GAFchromic films. Finally, the fD detector was used to fully characterize the 7 and 9 MeV UH‐DPP electron beams delivered by the ElectronFlash linac. In particular, PDDs, beamAbstract: Purpose: A diamond detector prototype was recently proposed by Marinelli et al. (Medical Physics 2022, https://doi.org/10.1002/mp.15473 ) for applications in ultrahigh‐dose‐per‐pulse (UH‐DPP) and ultrahigh‐dose‐rate (UH‐DR) beams, as used in FLASH radiotherapy (FLASH‐RT). In the present study, such so‐called flashDiamond (fD) was investigated from the dosimetric point of view, under pulsed electron beam irradiation. It was then used for the commissioning of an ElectronFlash linac (SIT S.p.A., Italy) both in conventional and UH‐DPP modalities. Methods: Detector calibration was performed in reference conditions, under 60 Co and electron beam irradiation. Its response linearity was investigated in UH‐DPP conditions. For this purpose, the DPP was varied in the 1.2–11.9 Gy range, by changing either the beam applicator or the pulse duration from 1 to 4 μs. Dosimetric validation of the fD detector prototype was then performed in conventional modality, by measuring percentage depth dose (PDD) curves, beam profiles, and output factors (OFs). All such measurements were carried out in a motorized water phantom. The obtained results were compared with the ones from commercially available dosimeters, namely, a microDiamond, an Advanced Markus ionization chamber, a silicon diode detector, and EBT‐XD GAFchromic films. Finally, the fD detector was used to fully characterize the 7 and 9 MeV UH‐DPP electron beams delivered by the ElectronFlash linac. In particular, PDDs, beam profiles, and OFs were measured, for both energies and all the applicators, and compared with the ones from EBT‐XD films irradiated in the same experimental conditions. Results: The fD calibration coefficient resulted to be independent from the investigated beam qualities. The detector response was found to be linear in the whole investigated DPP range. A very good agreement was observed among PDDs, beam profiles, and OFs measured by the fD prototype and reference detectors, both in conventional and UH‐DPP irradiation modalities. Conclusions: The fD detector prototype was validated from the dosimetric point of view against several commercial dosimeters in conventional beams. It was proved to be suitable in UH‐DPP and UH‐DR conditions, for which no other commercial real‐time active detector is available to date. It was shown to be a very useful tool to perform fast and reproducible beam characterizations in standard clinical motorized water phantom setups. All of the previously mentioned demonstrate the suitability of the proposed detector for the commissioning of UH‐DR linac beams for preclinical FLASH‐RT applications. … (more)
- Is Part Of:
- Medical physics. Volume 49:Issue 8(2022)
- Journal:
- Medical physics
- Issue:
- Volume 49:Issue 8(2022)
- Issue Display:
- Volume 49, Issue 8 (2022)
- Year:
- 2022
- Volume:
- 49
- Issue:
- 8
- Issue Sort Value:
- 2022-0049-0008-0000
- Page Start:
- 5513
- Page End:
- 5522
- Publication Date:
- 2022-06-16
- Subjects:
- diamond detector -- dosimetry -- FLASH radiotherapy
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.15782 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23103.xml