Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography. Issue 1 (29th October 2020)
- Record Type:
- Journal Article
- Title:
- Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography. Issue 1 (29th October 2020)
- Main Title:
- Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography
- Authors:
- Greve, Tobias
Sollmann, Nico
Hock, Andreas
Zimmer, Claus
Kirschke, Jan S. - Abstract:
- ABSTRACT: BACKGROUND AND PURPOSE: Intracranial vessel imaging by time‐of‐flight MR angiography (TOF‐MRA) is one of the most frequently performed investigations in clinical neuroradiology. Particularly in the acute setting, fast imaging is needed for diagnostics, with a sequence ideally depicting even small vessels. The purpose of this study was to compare image and diagnostic quality of a novel ultrashort TOF‐MRA sequence accelerated by spiral imaging (TOF‐Spiral‐short) to a standard TOF‐MRA sequence accelerated by compressed sensing (TOF‐CS) and to CT angiography (CTA). METHODS: Forty‐one patients (36.6% showing vessel pathologies) who had undergone TOF‐CS (acquisition duration: 4 minutes 8 seconds), TOF‐Spiral‐short (acquisition duration: 51 seconds; spiral imaging [accelerating factor 1.3], decreased field of view [accelerating factor 1.2], and increased voxel size [accelerating factor 3.3]), and CTA were retrospectively evaluated. Assessment of image quality, diagnostic confidence, and quantification of stenosis or aneurysm diameter were performed by two readers. RESULTS: Image quality at the skull base was slightly reduced with TOF‐Spiral‐short compared to CTA and TOF‐CS ( P < .05). Delineation of small intracranial vessels was improved by TOF‐Spiral‐short compared to CTA ( P < .0001). In TOF‐Spiral‐short, diagnostic confidence was not reduced compared to TOF‐CS in patients with vessel pathologies. We observed no significant difference in quantitative pathologyABSTRACT: BACKGROUND AND PURPOSE: Intracranial vessel imaging by time‐of‐flight MR angiography (TOF‐MRA) is one of the most frequently performed investigations in clinical neuroradiology. Particularly in the acute setting, fast imaging is needed for diagnostics, with a sequence ideally depicting even small vessels. The purpose of this study was to compare image and diagnostic quality of a novel ultrashort TOF‐MRA sequence accelerated by spiral imaging (TOF‐Spiral‐short) to a standard TOF‐MRA sequence accelerated by compressed sensing (TOF‐CS) and to CT angiography (CTA). METHODS: Forty‐one patients (36.6% showing vessel pathologies) who had undergone TOF‐CS (acquisition duration: 4 minutes 8 seconds), TOF‐Spiral‐short (acquisition duration: 51 seconds; spiral imaging [accelerating factor 1.3], decreased field of view [accelerating factor 1.2], and increased voxel size [accelerating factor 3.3]), and CTA were retrospectively evaluated. Assessment of image quality, diagnostic confidence, and quantification of stenosis or aneurysm diameter were performed by two readers. RESULTS: Image quality at the skull base was slightly reduced with TOF‐Spiral‐short compared to CTA and TOF‐CS ( P < .05). Delineation of small intracranial vessels was improved by TOF‐Spiral‐short compared to CTA ( P < .0001). In TOF‐Spiral‐short, diagnostic confidence was not reduced compared to TOF‐CS in patients with vessel pathologies. We observed no significant difference in quantitative pathology assessment between TOF‐Spiral‐short and the other two modalities. TOF‐Spiral‐short enabled the correct identification of all vessel pathologies. CONCLUSIONS: Accelerating TOF‐MRA of brain‐feeding arteries by a novel ultrashort spiral imaging sequence shows adequate image quality and sufficient diagnostic performance. Thus, TOF‐Spiral‐short holds potential for fast and reliable diagnostics of vessel pathologies, particularly in the acute setting. … (more)
- Is Part Of:
- Journal of neuroimaging. Volume 31:Issue 1(2021)
- Journal:
- Journal of neuroimaging
- Issue:
- Volume 31:Issue 1(2021)
- Issue Display:
- Volume 31, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 31
- Issue:
- 1
- Issue Sort Value:
- 2021-0031-0001-0000
- Page Start:
- 45
- Page End:
- 56
- Publication Date:
- 2020-10-29
- Subjects:
- Aneurysm -- cerebral arteries -- magnetic resonance angiography -- parallel imaging -- stroke
Diagnostic imaging -- Periodicals
Nervous system -- Diseases -- Diagnosis -- Periodicals
Imagerie pour le diagnostic -- Périodiques
Système nerveux -- Maladies -- Diagnostic -- Périodiques
Imagerie médicale
Neuroimagerie
Neurologie
Système nerveux
Périodique électronique (Descripteur de forme)
Ressource Internet (Descripteur de forme)
616.804754 - Journal URLs:
- http://jon.sagepub.com/ ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1552-6569 ↗
http://www.ingentaconnect.com/content/bpl/jon ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/jon.12791 ↗
- Languages:
- English
- ISSNs:
- 1051-2284
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5021.548000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23101.xml