Symmetric calibration method of pendulous Integrating Gyroscopic accelerometer on centrifuge. (15th August 2022)
- Record Type:
- Journal Article
- Title:
- Symmetric calibration method of pendulous Integrating Gyroscopic accelerometer on centrifuge. (15th August 2022)
- Main Title:
- Symmetric calibration method of pendulous Integrating Gyroscopic accelerometer on centrifuge
- Authors:
- Sun, C.
Ren, S.Q.
Cao, J.
Huo, R. - Abstract:
- Highlights: The closure errors in the complete error calibration model of PIGA significantly affect the calibration accuracy of PIGA. The proposed symmetric position calibration method can suppress the influence of the disk centrifuge errors and closure errors. The main error sources such as misalignments and installation errors could be automatically avoided by the symmetric position test. The integer periods sampling of PIGA precession can restrain the sampling error and significantly improve the output accuracy of PIGA. Abstract: With increasing requirements for the accuracy of Pendulous Integrating Gyroscopic Accelerometer (PIGA), developing effective methods that can accurately calibrate nonlinear error parameters of PIGA is a necessity. In this paper, the symmetric position calibration method is proposed to calibrate the main nonlinear error coefficients of PIGA within integer precession periods on the centrifuge. Firstly, coordinate systems are established, and specific forces, as well as angular velocities of PIGA, are deduced. Then, the complete error calibration model of PIGA, including the high-order error terms, is established. Calibration methods in symmetric positions are proposed, while the closure errors are restrained by the reasonable design of the test time. The centrifuge experiments show that the influence of misalignments could be restrained by the proposed calibration method, but the angular velocity of the main axis should be monitored. Thus, theHighlights: The closure errors in the complete error calibration model of PIGA significantly affect the calibration accuracy of PIGA. The proposed symmetric position calibration method can suppress the influence of the disk centrifuge errors and closure errors. The main error sources such as misalignments and installation errors could be automatically avoided by the symmetric position test. The integer periods sampling of PIGA precession can restrain the sampling error and significantly improve the output accuracy of PIGA. Abstract: With increasing requirements for the accuracy of Pendulous Integrating Gyroscopic Accelerometer (PIGA), developing effective methods that can accurately calibrate nonlinear error parameters of PIGA is a necessity. In this paper, the symmetric position calibration method is proposed to calibrate the main nonlinear error coefficients of PIGA within integer precession periods on the centrifuge. Firstly, coordinate systems are established, and specific forces, as well as angular velocities of PIGA, are deduced. Then, the complete error calibration model of PIGA, including the high-order error terms, is established. Calibration methods in symmetric positions are proposed, while the closure errors are restrained by the reasonable design of the test time. The centrifuge experiments show that the influence of misalignments could be restrained by the proposed calibration method, but the angular velocity of the main axis should be monitored. Thus, the impact of centrifuge errors on the measurement accuracy of PIGA is effectively suppressed, thereby reducing output uncertainty of PIGA to 10 −6 rad/s. The simulation results show that the order of calibration uncertainty of PIGA's second-order error parameter is decreased from 10 −6 to 10 −7 and the order of calibration uncertainty of other nonlinear error term coefficients is lowered to less than 10 −6 . Finally, the calibration accuracy of PIGA reaches 1 × 10 −7 g/g. … (more)
- Is Part Of:
- Measurement. Volume 200(2022)
- Journal:
- Measurement
- Issue:
- Volume 200(2022)
- Issue Display:
- Volume 200, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 200
- Issue:
- 2022
- Issue Sort Value:
- 2022-0200-2022-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-08-15
- Subjects:
- PIGA -- Nonlinear error term -- Calibration -- Error analysis -- Symmetric position -- Centrifuge
Weights and measures -- Periodicals
Measurement -- Periodicals
Measurement
Weights and measures
Periodicals
530.8 - Journal URLs:
- http://www.sciencedirect.com/science/journal/02632241 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.measurement.2022.111685 ↗
- Languages:
- English
- ISSNs:
- 0263-2241
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5413.544700
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23048.xml