Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans. Issue 8 (22nd June 2022)
- Record Type:
- Journal Article
- Title:
- Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans. Issue 8 (22nd June 2022)
- Main Title:
- Impact of systematic MLC positional uncertainties on the quality of single‐isocenter multi‐target VMAT‐SRS treatment plans
- Authors:
- Prentou, Georgia
Pappas, Eleftherios P
Prentou, Eleni
Yakoumakis, Nikolaos
Paraskevopoulou, Chryssa
Koutsouveli, Efi
Pantelis, Evaggelos
Papagiannis, Panagiotis
Karaiskos, Pantelis - Abstract:
- Abstract: Purpose: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub‐optimal MLC modeling) on the quality of intracranial single‐isocenter multi‐target VMAT‐SRS treatment plans. An estimation of appropriate tolerance levels is attempted. Methods: Five patients, with three to four metastases and at least one target lying in close proximity to organs‐at‐risk (OARs) were included in this study. A single‐isocenter multi‐arc VMAT plan per patient was prepared, which served as the reference for dosimetric impact evaluation. A range of leaf offsets was introduced (±0.03 mm up to ±0.30 mm defined at the MLC plane) to both leaf banks, by varying the leaf offset MLC modeling parameter in Monaco for all the prepared plans, in order to simulate projected leaf offsets of ±0.09 mm up to ±0.94 mm at the isocenter plane, respectively. For all offsets simulated and cases studied, dose distributions were re‐calculated and compared with the corresponding reference ones. An experimental dosimetric procedure using the SRS mapCHECK diode array was also performed to support the simulation study results and investigate its suitability to detect small systematic leaf positional errors. Results: Projected leaf offsets of ±0.09 mm were well‐tolerated with respect to both target dosimetry and OAR‐sparing. A linear relationship was found between D 95% percentage change and projected leaf offset (slope: 12%/mm). Impact of projected offset onAbstract: Purpose: To study the impact of systematic MLC leaf positional uncertainties (stemming from mechanical inaccuracies or sub‐optimal MLC modeling) on the quality of intracranial single‐isocenter multi‐target VMAT‐SRS treatment plans. An estimation of appropriate tolerance levels is attempted. Methods: Five patients, with three to four metastases and at least one target lying in close proximity to organs‐at‐risk (OARs) were included in this study. A single‐isocenter multi‐arc VMAT plan per patient was prepared, which served as the reference for dosimetric impact evaluation. A range of leaf offsets was introduced (±0.03 mm up to ±0.30 mm defined at the MLC plane) to both leaf banks, by varying the leaf offset MLC modeling parameter in Monaco for all the prepared plans, in order to simulate projected leaf offsets of ±0.09 mm up to ±0.94 mm at the isocenter plane, respectively. For all offsets simulated and cases studied, dose distributions were re‐calculated and compared with the corresponding reference ones. An experimental dosimetric procedure using the SRS mapCHECK diode array was also performed to support the simulation study results and investigate its suitability to detect small systematic leaf positional errors. Results: Projected leaf offsets of ±0.09 mm were well‐tolerated with respect to both target dosimetry and OAR‐sparing. A linear relationship was found between D 95% percentage change and projected leaf offset (slope: 12%/mm). Impact of projected offset on target dosimetry was strongly associated with target volume. In two cases, plans that could be considered potentially clinically unacceptable (i.e., clinical dose constraint violation) were obtained even for projected offsets as small as 0.19 mm. The performed experimental dosimetry check can detect potential small systematic leaf errors. Conclusions: Plan quality indices and dose–volume metrics are very sensitive to systematic sub‐millimeter leaf positional inaccuracies, projected at the isocenter plane. Acceptable and tolerance levels in systematic MLC uncertainties need to be tailored to VMAT‐SRS spatial and dosimetric accuracy requirements. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 23:Issue 8(2022)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 23:Issue 8(2022)
- Issue Display:
- Volume 23, Issue 8 (2022)
- Year:
- 2022
- Volume:
- 23
- Issue:
- 8
- Issue Sort Value:
- 2022-0023-0008-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-06-22
- Subjects:
- brain metastases -- leaf offset -- leaf positional uncertainty -- MLC -- single isocenter -- spatial accuracy -- SRS -- stereotactic radiosurgery -- VMAT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13708 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 23006.xml