Impact of Patient Interface Diameter and Vacuum Level on Suction Stability Using a Flat Applanating Interface for Femtosecond Laser-Assisted LASIK. (2nd July 2020)
- Record Type:
- Journal Article
- Title:
- Impact of Patient Interface Diameter and Vacuum Level on Suction Stability Using a Flat Applanating Interface for Femtosecond Laser-Assisted LASIK. (2nd July 2020)
- Main Title:
- Impact of Patient Interface Diameter and Vacuum Level on Suction Stability Using a Flat Applanating Interface for Femtosecond Laser-Assisted LASIK
- Authors:
- Mariacher, Siegfried
Mariacher, Martina
Seuthe, Anna Maria
Szurman, Peter
Seitz, Berthold
Boden, Karl Thomas - Abstract:
- ABSTRACT: Purpose: To investigate the impact of different patient interface (PI) diameters and different vacuum levels on the suction stability during vacuum application for femtosecond laser-assisted LASIK. Methods: Break-away forces as a marker for suction stability were measured by an automated test bench set-up. The test bench was based on a customized stamp connected to a digital load cell. Three flat applanating PI with different diameters (8.5, 9.5 and 10.0 mm) and four different vacuum levels (500, 600, 700 and 800 mbar) were investigated using the FEMTO LDV Z8 (Ziemer Ophthalmic Systems AG, Switzerland). Results: Mean break-away force was 5.23N (±0.99N) using the 8.5 mm PI and 500 mbar vacuum, 8.18N (±1.39N) using the 8.5 mm PI and 800 mbar, 3.37N (±0.56N) using the 10.0 mm PI and 500 mbar, and 6.14N (±0.68N) using the 10.0 mm PI and 800 mbar vacuum (p < .001; CI95%). Increasing the PI diameter from 8.5 to 10.0 mm resulted in a 28.89% (+1.97 ± 1.02N) lower break-away force (p < .001) when using default vacuum settings (700 mbar) compared to increasing vacuum from 500 to 800 mbar, which resulted in a 60.37% (+1.95 ± 1.40N) higher mean break away force (p < .001). Conclusion: The vacuum level and the diameter of the PI showed significant impact on suction stability measured as break-away force during flat applanating docking for corneal and refractive surgery. Break-away forces were inversely related to the PI diameter and directly to the vacuum level. Increasing theABSTRACT: Purpose: To investigate the impact of different patient interface (PI) diameters and different vacuum levels on the suction stability during vacuum application for femtosecond laser-assisted LASIK. Methods: Break-away forces as a marker for suction stability were measured by an automated test bench set-up. The test bench was based on a customized stamp connected to a digital load cell. Three flat applanating PI with different diameters (8.5, 9.5 and 10.0 mm) and four different vacuum levels (500, 600, 700 and 800 mbar) were investigated using the FEMTO LDV Z8 (Ziemer Ophthalmic Systems AG, Switzerland). Results: Mean break-away force was 5.23N (±0.99N) using the 8.5 mm PI and 500 mbar vacuum, 8.18N (±1.39N) using the 8.5 mm PI and 800 mbar, 3.37N (±0.56N) using the 10.0 mm PI and 500 mbar, and 6.14N (±0.68N) using the 10.0 mm PI and 800 mbar vacuum (p < .001; CI95%). Increasing the PI diameter from 8.5 to 10.0 mm resulted in a 28.89% (+1.97 ± 1.02N) lower break-away force (p < .001) when using default vacuum settings (700 mbar) compared to increasing vacuum from 500 to 800 mbar, which resulted in a 60.37% (+1.95 ± 1.40N) higher mean break away force (p < .001). Conclusion: The vacuum level and the diameter of the PI showed significant impact on suction stability measured as break-away force during flat applanating docking for corneal and refractive surgery. Break-away forces were inversely related to the PI diameter and directly to the vacuum level. Increasing the PI diameter by one step or reducing vacuum by 100 mbar resulted in a comparable decrease of break-away forces. Therefore, the surgeon could potentially maintain stabile suction by increasing vacuum when choosing a larger PI diameter to obtain a larger treatment zone. Furthermore, reduction of the PI diameter and/or increasing vacuum could help improving suction stability in situations of poor suction and in situations with increased risk of suction loss during LASIK. … (more)
- Is Part Of:
- Current eye research. Volume 45:Number 7(2020)
- Journal:
- Current eye research
- Issue:
- Volume 45:Number 7(2020)
- Issue Display:
- Volume 45, Issue 7 (2020)
- Year:
- 2020
- Volume:
- 45
- Issue:
- 7
- Issue Sort Value:
- 2020-0045-0007-0000
- Page Start:
- 789
- Page End:
- 796
- Publication Date:
- 2020-07-02
- Subjects:
- Patient interface -- vacuum -- suction -- applanation -- ziemer femto ldv z8
Ophthalmology -- Periodicals
Eye -- Diseases -- Periodicals
Ophthalmology -- Periodicals
573.88 - Journal URLs:
- http://informahealthcare.com/journal/cey ↗
http://www.tandfonline.com/toc/icey20/current ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/02713683.2019.1701688 ↗
- Languages:
- English
- ISSNs:
- 0271-3683
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3496.570000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22948.xml