The effect of clinically recommended Evolut sizes on anchorage forces after BASILICA. (30th March 2021)
- Record Type:
- Journal Article
- Title:
- The effect of clinically recommended Evolut sizes on anchorage forces after BASILICA. (30th March 2021)
- Main Title:
- The effect of clinically recommended Evolut sizes on anchorage forces after BASILICA
- Authors:
- Yaakobovich, Halit
Plitman Mayo, Romina
Zaretsky, Uri
Finkelstein, Ariel
Weiss, Dar
Marom, Gil - Abstract:
- Abstract: Coronary artery obstruction (CAO), a fatal complication of transcatheter aortic valve replacement (TAVR), is commonly found after Valve-in-Valve implantation inside a degenerated bioprosthetic valve. Leaflet laceration (BASILICA technique) has been proposed to prevent CAO and to potentially reduce the risk of leaflet thrombosis. We have previously demonstrated that this technique can reduce the anchorage forces of the TAVR device, which may lead to future complications. In this short communication, we hypothesize that the anchorage force reduction can be minimized by implanting a TAVR with a larger diameter, if two sizes are clinically recommended. We evaluated this hypothesis by employing finite element models of the deployments of the Evolut 26 and 29 mm inside a 27 mm Mitroflow valve, with and without leaflet lacerations. The results show that a laceration substantially decreases the contact area between the Evolut stent and the Mitroflow valve. The larger Evolut has a larger contact area and stronger anchorage forces. Additionally, the anchorage forces are less sensitive to additional lacerations in the larger Evolut (29 case). The results suggest that a larger self-expending device can ensure stronger anchorage and can lower the risk of possible migration, when TAVR is performed in a lacerated bioprosthesis.
- Is Part Of:
- Journal of biomechanics. Volume 118(2021)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 118(2021)
- Issue Display:
- Volume 118, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 118
- Issue:
- 2021
- Issue Sort Value:
- 2021-0118-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-03-30
- Subjects:
- TAVR -- BASILICA -- Finite element analysis
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2021.110303 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22878.xml