Feasibility of triple gamma ray imaging of 10C for range verification in ion therapy. (21st August 2022)
- Record Type:
- Journal Article
- Title:
- Feasibility of triple gamma ray imaging of 10C for range verification in ion therapy. (21st August 2022)
- Main Title:
- Feasibility of triple gamma ray imaging of 10C for range verification in ion therapy
- Authors:
- Mohammadi, Akram
Tashima, Hideaki
Takyu, Sodai
Iwao, Yuma
Akamatsu, Go
Kang, Han Gyu
Obata, Fujino
Nishikido, Fumihiko
Parodi, Katia
Yamaya, Taiga - Abstract:
- Abstract: Objective. In carbon ion therapy, the visualization of the range of incident particles in a patient body is important for treatment verification. In-beam positron emission tomography (PET) imaging is one of the methods to verify the treatment in ion therapy due to the high quality of PET images. We have shown the feasibility of in-beam PET imaging of radioactive 15 O and 11 C ion beams for range verification using our OpenPET system. Recently, we developed a whole gamma imager (WGI) that can simultaneously work as PET, single gamma ray and triple gamma ray imaging. The WGI has high potential to detect the location of 10 C, which emits positrons with a simultaneous gamma ray of 718 keV, within the patient's body during ion therapy. Approach. In this work, we focus on investigating the performance of WGI for 10 C imaging and its feasibility for range verification in carbon ion therapy. First, the performance of the WGI was studied to image a 10 C point source using the Geant4 toolkit. Then, the feasibility of WGI was investigated for an irradiated polymethyl methacrylate (PMMA) phantom with a 10 C ion beam at the carbon therapy facility of the Heavy Ion Medical Accelerator in Chiba. Main results. The average spatial resolution and sensitivity for the simulated 10 C point source at the centre of the field of view were 5.5 mm FWHM and 0.010%, respectively. The depth dose of the 10 C ion beam was measured, and the triple gamma image of 10 C nuclides for an irradiatedAbstract: Objective. In carbon ion therapy, the visualization of the range of incident particles in a patient body is important for treatment verification. In-beam positron emission tomography (PET) imaging is one of the methods to verify the treatment in ion therapy due to the high quality of PET images. We have shown the feasibility of in-beam PET imaging of radioactive 15 O and 11 C ion beams for range verification using our OpenPET system. Recently, we developed a whole gamma imager (WGI) that can simultaneously work as PET, single gamma ray and triple gamma ray imaging. The WGI has high potential to detect the location of 10 C, which emits positrons with a simultaneous gamma ray of 718 keV, within the patient's body during ion therapy. Approach. In this work, we focus on investigating the performance of WGI for 10 C imaging and its feasibility for range verification in carbon ion therapy. First, the performance of the WGI was studied to image a 10 C point source using the Geant4 toolkit. Then, the feasibility of WGI was investigated for an irradiated polymethyl methacrylate (PMMA) phantom with a 10 C ion beam at the carbon therapy facility of the Heavy Ion Medical Accelerator in Chiba. Main results. The average spatial resolution and sensitivity for the simulated 10 C point source at the centre of the field of view were 5.5 mm FWHM and 0.010%, respectively. The depth dose of the 10 C ion beam was measured, and the triple gamma image of 10 C nuclides for an irradiated PMMA phantom was obtained by applying a simple back projection to the detected triple gammas. Significance. The shift between Bragg peak position and position of the peak of the triple gamma image in an irradiated PMMA phantom was 2.8 ± 0.8 mm, which demonstrates the capability of triple gamma imaging using WGI for range verification of 10 C ion beams. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 67:Number 16(2022)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 67:Number 16(2022)
- Issue Display:
- Volume 67, Issue 16 (2022)
- Year:
- 2022
- Volume:
- 67
- Issue:
- 16
- Issue Sort Value:
- 2022-0067-0016-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-08-21
- Subjects:
- triple gamma imaging -- 10C nuclide -- 10C ion beam -- range verification -- visualization of 10C
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ac826a ↗