Novel Method for Implantation of Balloon Expandable Transcatheter Aortic Valve Replacement to Reduce Pacemaker Rate—Line of Lucency Method. Issue 5 (2nd September 2020)
- Record Type:
- Journal Article
- Title:
- Novel Method for Implantation of Balloon Expandable Transcatheter Aortic Valve Replacement to Reduce Pacemaker Rate—Line of Lucency Method. Issue 5 (2nd September 2020)
- Main Title:
- Novel Method for Implantation of Balloon Expandable Transcatheter Aortic Valve Replacement to Reduce Pacemaker Rate—Line of Lucency Method
- Authors:
- Ramanathan, P. Kasi
Nazir, Salik
Elzanaty, Ahmed M.
Nesheiwat, Zeid
Mahmood, Muhammad
Rachwal, William
Riordan, Christopher
Letcher, John
Yenrick, Kellie
Boonie, Erica
Moront, Michael G.
Redfern, Roberta E.
Crescenzo, Donald - Abstract:
- ABSTRACT: Background: The risk of permanent pacemaker implantation (PPM) following transcatheter aortic valve replacement (TAVR) remains high. Valve implantation depth is an independent predictor of PPM. This study describes the rate of PPM following a novel method for valve positioning during TAVR. Methods: A retrospective cohort study involved patients undergoing TAVR at our institute between January 2019 and January 2020. Baseline characteristics, procedural data, and clinical outcomes were collected and analyzed. All patients were implanted with the Edwards SAPIEN 3 or SAPIEN 3 Ultra using the radiolucent line on the stented valve, rather than the central marker, to guide positioning. We describe this as the line of lucency method. Results: In total, 50 patients (96% SAPIEN 3, and 4% SAPIEN 3 Ultra) were included. The median age was 81.5 years, and 30 (60%) patients had existing conduction abnormalities: first degree atrio-ventricular block (18%), right bundle branch block (12%), left bundle branch block (10.2%), and left anterior fascicular block (16%). The modified technique was successful in all cases. No patient required PPM within 30 days. Two patients (4%) had moderate aortic regurgitation on echocardiogram at 30 days; 3 (6%) developed new left bundle branch block. There was no instance of valve embolization, coronary obstruction, or postprocedural death. Conclusion: A novel technique utilizing the radiolucent line on the bottom of the stented valve during TAVRABSTRACT: Background: The risk of permanent pacemaker implantation (PPM) following transcatheter aortic valve replacement (TAVR) remains high. Valve implantation depth is an independent predictor of PPM. This study describes the rate of PPM following a novel method for valve positioning during TAVR. Methods: A retrospective cohort study involved patients undergoing TAVR at our institute between January 2019 and January 2020. Baseline characteristics, procedural data, and clinical outcomes were collected and analyzed. All patients were implanted with the Edwards SAPIEN 3 or SAPIEN 3 Ultra using the radiolucent line on the stented valve, rather than the central marker, to guide positioning. We describe this as the line of lucency method. Results: In total, 50 patients (96% SAPIEN 3, and 4% SAPIEN 3 Ultra) were included. The median age was 81.5 years, and 30 (60%) patients had existing conduction abnormalities: first degree atrio-ventricular block (18%), right bundle branch block (12%), left bundle branch block (10.2%), and left anterior fascicular block (16%). The modified technique was successful in all cases. No patient required PPM within 30 days. Two patients (4%) had moderate aortic regurgitation on echocardiogram at 30 days; 3 (6%) developed new left bundle branch block. There was no instance of valve embolization, coronary obstruction, or postprocedural death. Conclusion: A novel technique utilizing the radiolucent line on the bottom of the stented valve during TAVR provided reproducibly high positioning with excellent clinical results. Optimal position of the valve may prevent need for subsequent PPM. … (more)
- Is Part Of:
- Structural heart. Volume 4:Issue 5(2020)
- Journal:
- Structural heart
- Issue:
- Volume 4:Issue 5(2020)
- Issue Display:
- Volume 4, Issue 5 (2020)
- Year:
- 2020
- Volume:
- 4
- Issue:
- 5
- Issue Sort Value:
- 2020-0004-0005-0000
- Page Start:
- 427
- Page End:
- 432
- Publication Date:
- 2020-09-02
- Subjects:
- Balloon expandable bioprosthesis -- SAPIEN 3 -- TAVR -- high implantation -- permanent pacemaker
Heart -- Diseases -- Periodicals
Congenital heart disease -- Periodicals
Cardiovascular system -- Diseases -- Periodicals
Cardiovascular Diseases
Cardiovascular system -- Diseases
Congenital heart disease
Heart -- Diseases
Periodicals
616.12 - Journal URLs:
- http://www.tandfonline.com/loi/ushj20 ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/24748706.2020.1813355 ↗
- Languages:
- English
- ISSNs:
- 2474-8706
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22727.xml