Influence of the Reamer-Irrigator-Aspirator diameter on femoral bone strength and amount of harvested bone graft – a biomechanical cadaveric study. Issue 12 (December 2020)
- Record Type:
- Journal Article
- Title:
- Influence of the Reamer-Irrigator-Aspirator diameter on femoral bone strength and amount of harvested bone graft – a biomechanical cadaveric study. Issue 12 (December 2020)
- Main Title:
- Influence of the Reamer-Irrigator-Aspirator diameter on femoral bone strength and amount of harvested bone graft – a biomechanical cadaveric study
- Authors:
- Schmitz, Nina
Gehweiler, Dominic
Wähnert, Dirk
Zderic, Ivan
Grünwald, Leonard
Richards, Geoff
Gueorguiev, Boyko
Raschke, Michael - Abstract:
- Highlights: RIA diameter of up to 2.5 mm larger than the isthmus endosteal diameter seems to be safe whereas 4.0 mm leads to a reduction of femoral bone strength. The bone graft amount harvested with RIA increased with larger diameter being significantly bigger at plus 4.0 mm diameter than plus 1.5 mm or 2.5 mm. For determination of the RIA reaming diameter conventional X-rays appears to be as precise as a CT scan. Abstract: Background: Treatment of large bone defects is still related to unsolved problems in orthopaedic trauma surgery. Minimally invasive intramedullary reaming with the use of the Reamer-Irrigator-Aspirator (RIA) device allows autograft harvesting of large bone graft amounts from the medullary canal of the femur. The aim of this study was to investigate the influence of RIA diameter on femoral bone strength and amount of harvested bone graft in a human cadaveric model. Methods: Forty-five pairs human cadaveric femora were randomized to 3 paired groups with 15 pairs each. One femur of each pair was reamed with RIA at a diameter of either 1.5 mm (group 1), 2.5 mm (group 2) or 4.0 mm (group 3) larger than its isthmus, whereas its contralateral femur was left intact without reaming. The amount of harvested bone graft was determined for each specimen and all femora were destructively tested in internal rotation under 750 N axial compression to calculate their torsional stiffness and torque at failure. Results: Significant reduction in torsional stiffness wasHighlights: RIA diameter of up to 2.5 mm larger than the isthmus endosteal diameter seems to be safe whereas 4.0 mm leads to a reduction of femoral bone strength. The bone graft amount harvested with RIA increased with larger diameter being significantly bigger at plus 4.0 mm diameter than plus 1.5 mm or 2.5 mm. For determination of the RIA reaming diameter conventional X-rays appears to be as precise as a CT scan. Abstract: Background: Treatment of large bone defects is still related to unsolved problems in orthopaedic trauma surgery. Minimally invasive intramedullary reaming with the use of the Reamer-Irrigator-Aspirator (RIA) device allows autograft harvesting of large bone graft amounts from the medullary canal of the femur. The aim of this study was to investigate the influence of RIA diameter on femoral bone strength and amount of harvested bone graft in a human cadaveric model. Methods: Forty-five pairs human cadaveric femora were randomized to 3 paired groups with 15 pairs each. One femur of each pair was reamed with RIA at a diameter of either 1.5 mm (group 1), 2.5 mm (group 2) or 4.0 mm (group 3) larger than its isthmus, whereas its contralateral femur was left intact without reaming. The amount of harvested bone graft was determined for each specimen and all femora were destructively tested in internal rotation under 750 N axial compression to calculate their torsional stiffness and torque at failure. Results: Significant reduction in torsional stiffness was detected after reaming in group 3 ( p = 0.03) in contrast to groups 1 and 2 where no such significant reduction was observed ( p ≥ 0.34). Torque at failure was significantly reduced after reaming in all 3 groups ( p ≤ 0.04). Collected bone graft amount in group 3 was significantly bigger compared to groups 1 and 2 ( p ≤ 0.04). Conclusions: Reaming with RIA diameter of 4.0 mm larger than the isthmus of the femur seems to influence considerably its torsional stiffness, however, it allows harvesting of a significantly bigger bone graft amount. … (more)
- Is Part Of:
- Injury. Volume 51:Issue 12(2020)
- Journal:
- Injury
- Issue:
- Volume 51:Issue 12(2020)
- Issue Display:
- Volume 51, Issue 12 (2020)
- Year:
- 2020
- Volume:
- 51
- Issue:
- 12
- Issue Sort Value:
- 2020-0051-0012-0000
- Page Start:
- 2846
- Page End:
- 2850
- Publication Date:
- 2020-12
- Subjects:
- Reamer-Irrigator-Aspirator (RIA) Diameter -- Biomechanics -- Femoral Fracture -- Autologous Bone Graft Harvesting
Wounds and injuries -- Surgery -- Periodicals
Accidents -- Periodicals
Wounds and Injuries -- surgery -- Periodicals
Lésions et blessures -- Chirurgie -- Périodiques
Electronic journals
Electronic journals
617.1 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00201383 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00201383 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00201383 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.injury.2020.09.057 ↗
- Languages:
- English
- ISSNs:
- 0020-1383
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4514.400000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22645.xml