Development and evaluation of time‐resolved fluorescent immunochromatographic assay for quantitative detection of SARS‐CoV‐2 spike antigen. Issue 7 (12th June 2022)
- Record Type:
- Journal Article
- Title:
- Development and evaluation of time‐resolved fluorescent immunochromatographic assay for quantitative detection of SARS‐CoV‐2 spike antigen. Issue 7 (12th June 2022)
- Main Title:
- Development and evaluation of time‐resolved fluorescent immunochromatographic assay for quantitative detection of SARS‐CoV‐2 spike antigen
- Authors:
- Xu, Buzhou
Tang, Hao
Weng, Yiming
Jones, Valerie Sloane
Luo, Shuhong
Cho, Chih Yun
Lin, Yongping
Fang, Jianmin
Song, Xuedong
Huang, Ruo‐Pan - Abstract:
- Abstract: Background: The spread of COVID‐19 worldwide caused by the severe acute respiratory syndrome coronavirus 2 (SARS‐CoV‐2) has necessitated efficient, sensitive diagnostic methods to identify infected people. We report on the development of a rapid 15‐minute time‐resolved fluorescent (TRF) lateral flow immunochromatographic assay for the quantitative detection of the SARS‐CoV‐2 spike protein receptor‐binding domain (S1‐RBD). Objectives: Our objective was to develop an efficient method of detecting SARS‐CoV‐2 within 15 min of sample collection. Methods: We constructed and evaluated a portable, disposable lateral flow device, which detected the S1‐RBD protein directly in nasopharyngeal swab samples. The device emits a fluorescent signal in the presence of S1‐RBD, which can be captured by an automated TRF instrument. Results: The TRF lateral flow assay signal was linear from 0 to 20 ng/ml and demonstrated high accuracy and reproducibility. When evaluated with clinical nasopharyngeal swabs, the assay was performed at >80% sensitivity, >84% specificity, and > 82% accuracy for detection of the S1‐RBD antigen. Conclusion: The new S1‐RBD antigen test is a rapid (15 min), sensitive, and specific assay that requires minimal sample preparation. Critically, the assay correlated closely with PCR‐based methodology in nasopharyngeal swab samples, showing that the detected S1‐RBD antigen levels correlate with SARS‐CoV‐2 virus load. Therefore, the new TRF lateral flow test for S1‐RBDAbstract: Background: The spread of COVID‐19 worldwide caused by the severe acute respiratory syndrome coronavirus 2 (SARS‐CoV‐2) has necessitated efficient, sensitive diagnostic methods to identify infected people. We report on the development of a rapid 15‐minute time‐resolved fluorescent (TRF) lateral flow immunochromatographic assay for the quantitative detection of the SARS‐CoV‐2 spike protein receptor‐binding domain (S1‐RBD). Objectives: Our objective was to develop an efficient method of detecting SARS‐CoV‐2 within 15 min of sample collection. Methods: We constructed and evaluated a portable, disposable lateral flow device, which detected the S1‐RBD protein directly in nasopharyngeal swab samples. The device emits a fluorescent signal in the presence of S1‐RBD, which can be captured by an automated TRF instrument. Results: The TRF lateral flow assay signal was linear from 0 to 20 ng/ml and demonstrated high accuracy and reproducibility. When evaluated with clinical nasopharyngeal swabs, the assay was performed at >80% sensitivity, >84% specificity, and > 82% accuracy for detection of the S1‐RBD antigen. Conclusion: The new S1‐RBD antigen test is a rapid (15 min), sensitive, and specific assay that requires minimal sample preparation. Critically, the assay correlated closely with PCR‐based methodology in nasopharyngeal swab samples, showing that the detected S1‐RBD antigen levels correlate with SARS‐CoV‐2 virus load. Therefore, the new TRF lateral flow test for S1‐RBD has potential application in point‐of‐care settings. Abstract : We have developed a portable, disposable, and quantitative lateral flow device, which detects the SARS‐CoV‐2 Spike antigen in 15 min using time‐resolved fluorescence. Compared with RT‐PCR, this assay has a sensitivity of 80.3% and a specificity of 83.64% in a cohort of 293 patients. … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 36:Issue 7(2022)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 36:Issue 7(2022)
- Issue Display:
- Volume 36, Issue 7 (2022)
- Year:
- 2022
- Volume:
- 36
- Issue:
- 7
- Issue Sort Value:
- 2022-0036-0007-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-06-12
- Subjects:
- COVID‐19 -- RT‐PCR -- SARS‐CoV‐2 -- Spike protein -- time‐resolved fluorescence -- TRF lateral flow
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.24513 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22601.xml