Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile. Issue 7 (11th April 2022)
- Record Type:
- Journal Article
- Title:
- Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile. Issue 7 (11th April 2022)
- Main Title:
- Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Profile
- Authors:
- Carr, Madeline E.
Keenan, Kathryn E.
Rai, Robba
Boss, Michael A.
Metcalfe, Peter
Walker, Amy
Holloway, Lois - Abstract:
- Abstract: Purpose: To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice‐encoded imaging direction and the reference sample ADC value. Methods: ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements of ADC %bias, coefficients of variation for short‐/long‐term repeatability and precision (CVST /CVLT and CVP ), and b ‐value dependency (Dep b ) were calculated. The measurements were then assessed according to the Quantitative Imaging Biomarker Alliance (QIBA) Diffusion Profile specifications. Results: The average of all measurements over the year was within Profile recommended ranges. This included when testing was performed in different imaging directions, and on samples that had different ADC reference values (0.4–1.1 μm 2 /ms). Results in the axial plane for the central water vial included a bias of +0.05%, CVST /CVLT /CVP = 0.1%/ 0.9%/0.4% and Dep b = 0.4%. Conclusions: The technical performance of ADC on a radiotherapy dedicated MRI scanner over the course of 12 months was considered conformant to the QIBA Profile. Quantifying these metrics and factors that may affect the performance is essential in progressing the use of ADC clinically: ensuring that the observed change of ADCAbstract: Purpose: To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice‐encoded imaging direction and the reference sample ADC value. Methods: ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements of ADC %bias, coefficients of variation for short‐/long‐term repeatability and precision (CVST /CVLT and CVP ), and b ‐value dependency (Dep b ) were calculated. The measurements were then assessed according to the Quantitative Imaging Biomarker Alliance (QIBA) Diffusion Profile specifications. Results: The average of all measurements over the year was within Profile recommended ranges. This included when testing was performed in different imaging directions, and on samples that had different ADC reference values (0.4–1.1 μm 2 /ms). Results in the axial plane for the central water vial included a bias of +0.05%, CVST /CVLT /CVP = 0.1%/ 0.9%/0.4% and Dep b = 0.4%. Conclusions: The technical performance of ADC on a radiotherapy dedicated MRI scanner over the course of 12 months was considered conformant to the QIBA Profile. Quantifying these metrics and factors that may affect the performance is essential in progressing the use of ADC clinically: ensuring that the observed change of ADC in a tissue is due to a physiological response and not measurement variability. … (more)
- Is Part Of:
- Medical physics. Volume 49:Issue 7(2022)
- Journal:
- Medical physics
- Issue:
- Volume 49:Issue 7(2022)
- Issue Display:
- Volume 49, Issue 7 (2022)
- Year:
- 2022
- Volume:
- 49
- Issue:
- 7
- Issue Sort Value:
- 2022-0049-0007-0000
- Page Start:
- 4508
- Page End:
- 4517
- Publication Date:
- 2022-04-11
- Subjects:
- apparent diffusion coefficient -- diffusion‐weighted imaging -- quantitative imaging biomarker alliance -- quantitative magnetic resonance imaging -- reproducibility
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.15645 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22615.xml