Determination of the responsivity of a predictable quantum efficient detector over a wide spectral range based on a 3D model of charge carrier recombination losses. (1st August 2022)
- Record Type:
- Journal Article
- Title:
- Determination of the responsivity of a predictable quantum efficient detector over a wide spectral range based on a 3D model of charge carrier recombination losses. (1st August 2022)
- Main Title:
- Determination of the responsivity of a predictable quantum efficient detector over a wide spectral range based on a 3D model of charge carrier recombination losses
- Authors:
- Tran, Trinh
Porrovecchio, Geiland
Smid, Marek
Ikonen, Erkki
Dönsberg, Timo
Gran, Jarle - Abstract:
- Abstract: We present a method to determine the internal quantum deficiency (IQD) of a predictable quantum efficient detector (PQED) based on measured photocurrent dependence on bias voltage and a 3D simulation model of charge carrier recombination losses. The simulation model of silicon photodiodes includes wafer doping concentration, fixed charge of SiO2 layer, bulk lifetime of charge carriers and surface recombination velocity as the fitted parameters. With only one set of physical photodiode defining parameters, the simulation shows excellent agreement with experimental data at power levels from 100 μW to 1000 μW with variation in illumination beam size. We could also predict the dependence of IQD on bias voltage at the wavelength of 476 nm using photodiode parameters determined independently at 647 nm wavelength. The fitted values of doping concentration and fixed charge extracted from the simulation model are in close agreement with the expected parameter values determined earlier. At bias voltages larger than 5 V at the wavelength of 476 nm, the internal quantum efficiency of one of the tested PQEDs is measured to be 0.999 970 ± 0.000 027, where the relative expanded uncertainty of 0.000 027 is one of the lowest values ever achieved in spectral responsivity measurement of optical detectors.
- Is Part Of:
- Metrologia. Volume 59:Number 4(2022)
- Journal:
- Metrologia
- Issue:
- Volume 59:Number 4(2022)
- Issue Display:
- Volume 59, Issue 4 (2022)
- Year:
- 2022
- Volume:
- 59
- Issue:
- 4
- Issue Sort Value:
- 2022-0059-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-08-01
- Subjects:
- predictable quantum efficient detector -- simulation -- photodetector -- internal quantum deficiency
Weights and measures -- Periodicals
Weights and Measures -- Periodicals
530.805 - Journal URLs:
- http://iopscience.iop.org/0026-1394/ ↗
http://www.iop.org/ej/journal/0026-1394 ↗
http://www.iop.org/ ↗
http://www.ingentaconnect.com/content/bipm/met ↗ - DOI:
- 10.1088/1681-7575/ac604b ↗
- Languages:
- English
- ISSNs:
- 0026-1394
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 22575.xml