Independent recalculation outperforms traditional measurement‐based IMRT QA methods in detecting unacceptable plans. Issue 8 (23rd June 2019)
- Record Type:
- Journal Article
- Title:
- Independent recalculation outperforms traditional measurement‐based IMRT QA methods in detecting unacceptable plans. Issue 8 (23rd June 2019)
- Main Title:
- Independent recalculation outperforms traditional measurement‐based IMRT QA methods in detecting unacceptable plans
- Authors:
- Kry, Stephen F.
Glenn, Mallory C.
Peterson, Christine B.
Branco, Daniela
Mehrens, Hunter
Steinmann, Angela
Followill, David S. - Abstract:
- Abstract : Purpose: To evaluate the performance of an independent recalculation and compare it against current measurement‐based patient specific intensity‐modulated radiation therapy (IMRT) quality assurance (QA) in predicting unacceptable phantom results as measured by the Imaging and Radiation Oncology Core (IROC). Methods: When institutions irradiate the IROC head and neck IMRT phantom, they are also asked to submit their internal IMRT QA results. Separately from this, IROC has previously created reference beam models on the Mobius3D platform to independently recalculate phantom results based on the institution's DICOM plan data. The ability of the institutions' IMRT QA to predict the IROC phantom result was compared against the independent recalculation for 339 phantom results collected since 2012. This was done to determine the ability of these systems to detect failing phantom results (i.e., large errors) as well as poor phantom results (i.e., modest errors). Sensitivity and specificity were evaluated using common clinical thresholds, and receiver operator characteristic (ROC) curves were used to compare across different thresholds. Results: Overall, based on common clinical criteria, the independent recalculation was 12 times more sensitive at detecting unacceptable (failing) IROC phantom results than clinical measurement‐based IMRT QA. The recalculation was superior, in head‐to‐head comparison, to the EPID, ArcCheck, and MapCheck devices. The superiority of theAbstract : Purpose: To evaluate the performance of an independent recalculation and compare it against current measurement‐based patient specific intensity‐modulated radiation therapy (IMRT) quality assurance (QA) in predicting unacceptable phantom results as measured by the Imaging and Radiation Oncology Core (IROC). Methods: When institutions irradiate the IROC head and neck IMRT phantom, they are also asked to submit their internal IMRT QA results. Separately from this, IROC has previously created reference beam models on the Mobius3D platform to independently recalculate phantom results based on the institution's DICOM plan data. The ability of the institutions' IMRT QA to predict the IROC phantom result was compared against the independent recalculation for 339 phantom results collected since 2012. This was done to determine the ability of these systems to detect failing phantom results (i.e., large errors) as well as poor phantom results (i.e., modest errors). Sensitivity and specificity were evaluated using common clinical thresholds, and receiver operator characteristic (ROC) curves were used to compare across different thresholds. Results: Overall, based on common clinical criteria, the independent recalculation was 12 times more sensitive at detecting unacceptable (failing) IROC phantom results than clinical measurement‐based IMRT QA. The recalculation was superior, in head‐to‐head comparison, to the EPID, ArcCheck, and MapCheck devices. The superiority of the recalculation vs these array‐based measurements persisted under ROC analysis as the recalculation curve had a greater area under it and was always above that for these measurement devices. For detecting modest errors (poor phantom results rather than failing phantom results), neither the recalculation nor measurement‐based IMRT QA performed well. Conclusions: A simple recalculation outperformed current measurement‐based IMRT QA methods at detecting unacceptable plans. These findings highlight the value of an independent recalculation, and raise further questions about the current standard of measurement‐based IMRT QA. … (more)
- Is Part Of:
- Medical physics. Volume 46:Issue 8(2019)
- Journal:
- Medical physics
- Issue:
- Volume 46:Issue 8(2019)
- Issue Display:
- Volume 46, Issue 8 (2019)
- Year:
- 2019
- Volume:
- 46
- Issue:
- 8
- Issue Sort Value:
- 2019-0046-0008-0000
- Page Start:
- 3700
- Page End:
- 3708
- Publication Date:
- 2019-06-23
- Subjects:
- IMRT QA -- Mobius -- patient‐specific QA -- recalculation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.13638 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22507.xml