Bioprosthetic valve fracture during valve-in-valve transcatheter aortic valve replacement. Issue 3 (2nd July 2020)
- Record Type:
- Journal Article
- Title:
- Bioprosthetic valve fracture during valve-in-valve transcatheter aortic valve replacement. Issue 3 (2nd July 2020)
- Main Title:
- Bioprosthetic valve fracture during valve-in-valve transcatheter aortic valve replacement
- Authors:
- Hamandi, Mohanad
Nwafor, Ikenna
Hebeler, Katherine R.
Crawford, Alexander
Lanfear, Allison T.
Schaffer, Justin
Al-Azizi, Karim
Potluri, Srinivasa
Brinkman, William T.
Harrington, Katherine
Szerlip, Molly
DiMaio, J. Michael - Abstract:
- Abstract: Transcatheter aortic valve replacement (TAVR) valve-in-valve (VIV) therapy has been approved for select patients with surgically inoperable bioprosthetic valves that need replacement. Bioprosthetic valve fracturing (BVF) used in conjunction with VIV TAVR can reduce transvalvular gradients and increase the aortic valve area. Twelve patients who underwent BVF VIV TAVR at a single center were retrospectively analyzed. Measurements of hemodynamics and aortic valve area were performed at baseline, after VIV TAVR, after BVF, and at 30-day follow-up. The mean Society of Thoracic Surgeons Predicted Risk of Mortality score was 7.12 ± 3.5%, with 75% of patients deemed high risk by the heart team. Mean gradients decreased from 44 mm Hg to 15 mm Hg following VIV TAVR, and to 7 mm Hg following BVF. The mean aortic valve area increased from 0.6 cm 2 to 1 cm 2 following VIV TAVR, and to 1.3 cm 2 following BVF. There were no postoperative permanent pacemaker implantations or vascular complications, and at 30 days, only one patient had died. While we report intraoperative mortality, BVF with VIV TAVR can be performed to reduce transvalvular gradients and increase effective aortic valve area in high-surgical-risk patients with failed bioprosthetic valves.
- Is Part Of:
- Proceedings. Volume 33:Issue 3(2020)
- Journal:
- Proceedings
- Issue:
- Volume 33:Issue 3(2020)
- Issue Display:
- Volume 33, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 33
- Issue:
- 3
- Issue Sort Value:
- 2020-0033-0003-0000
- Page Start:
- 317
- Page End:
- 321
- Publication Date:
- 2020-07-02
- Subjects:
- Bioprosthetic valve fracturing -- transcatheter aortic valve replacement -- valve-in-valve
Medicine -- Periodicals
Medicine
Medicine -- Periodicals
Periodicals
Electronic journals
616.005 - Journal URLs:
- http://www.baylorhealth.edu/proceedings/default.htm ↗
https://www.tandfonline.com/loi/ubmc20 ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/08998280.2020.1732267 ↗
- Languages:
- English
- ISSNs:
- 0899-8280
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22370.xml