A low-cost geometry calibration procedure for a modular cone-beam X-ray CT system. Issue 3 (2nd July 2020)
- Record Type:
- Journal Article
- Title:
- A low-cost geometry calibration procedure for a modular cone-beam X-ray CT system. Issue 3 (2nd July 2020)
- Main Title:
- A low-cost geometry calibration procedure for a modular cone-beam X-ray CT system
- Authors:
- Nguyen, V.
De Beenhouwer, J.
Sanctorum, J. G.
Van Wassenbergh, S.
Bazrafkan, S.
Dirckx, J. J. J.
Sijbers, J. - Abstract:
- ABSTRACT: Accurate knowledge of the acquisition geometry of a CT scanning system is key for high quality tomographic imaging. Unfortunately, in modular X-ray CT setups, geometry misalignment occurs each time the setup is changed, which calls for an efficient calibration procedure to correct for geometric inaccuracies. Although many studies have been dealing with the calibration of X-ray CT systems, these are often specifically designed for one setup and/or expensive. In this work, we explore the possibilities of a low-cost, easy-to-build, and modular phantom, constructed from LEGO bricks, which serves as a structure to hold small metal beads, for geometric calibration of a tomographic X-ray system. By estimating the bead coordinates using deep learning, and minimizing center-to-center distances of the metal beads between measured and reference projection data, geometry parameters are derived. With simulated as well as real experiments, it is shown that the LEGO phantom can be used to accurately estimate the geometry of a modular X-ray CT system.
- Is Part Of:
- Nondestructive testing and evaluation. Volume 35:Issue 3(2020)
- Journal:
- Nondestructive testing and evaluation
- Issue:
- Volume 35:Issue 3(2020)
- Issue Display:
- Volume 35, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 35
- Issue:
- 3
- Issue Sort Value:
- 2020-0035-0003-0000
- Page Start:
- 252
- Page End:
- 265
- Publication Date:
- 2020-07-02
- Subjects:
- LEGO phantom -- geometry calibration -- cone-beam CT -- deep learning
Non-destructive testing -- Periodicals
620.112705 - Journal URLs:
- http://journalsonline.tandf.co.uk/app/home/journal.asp?wasp=23tyjmuxtj5xnmgunm13&referrer=parent&backto=searchpublicationsresults, 1, 1;homemain, 1, 1; ↗
http://www.tandfonline.com/toc/gnte20/current ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/10589759.2020.1774580 ↗
- Languages:
- English
- ISSNs:
- 1058-9759
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6117.044700
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 22360.xml