Comparison of scattered entrance skin dose burden in MSCT, CBCT, and X-ray for suspected scaphoid injury: Regional dose measurements in a phantom model. Issue 3 (August 2022)
- Record Type:
- Journal Article
- Title:
- Comparison of scattered entrance skin dose burden in MSCT, CBCT, and X-ray for suspected scaphoid injury: Regional dose measurements in a phantom model. Issue 3 (August 2022)
- Main Title:
- Comparison of scattered entrance skin dose burden in MSCT, CBCT, and X-ray for suspected scaphoid injury: Regional dose measurements in a phantom model
- Authors:
- Hughes, J.
Harris, M.
Snaith, B.
Benn, H. - Abstract:
- Abstract: Introduction: Scaphoid radiography has poor sensitivity for acute fracture detection and often requires repeat delayed imaging. Although magnetic resonance (MR) imaging is considered the gold standard, computed tomography (CT) is often used as an alternative due to ease of access. Cone-Beam CT (CBCT) offers equivalent diagnostic efficacy to Multi Slice CT (MSCT) at reduced dose. We aimed to establish the difference in scattered dose between modalities for scaphoid imaging. Methods: Anatomical regional entrance surface dose measurements were taken at 3 regions on an anthropomorphic torso phantom positioned as a patient to a wrist phantom undergoing scaphoid imaging for three modalities (CBCT, MSCT, four-view projection radiography). Exposure factors were based on audit of clinical exposures. Each dose measurement was repeated three times per anatomical region, modality, exposure setting and projection. Results: Under unpaired T-test CBCT gave significantly lower mean dose at the neck (1.64 vs 18 mGy), chest (2.78 vs 8.01) and abdomen (1.288 vs 2.93) than MSCT (p < .0001). However CBCT had significantly higher mean dose than four-view radiography at the neck, chest and abdomen (0.031, 0.035, and 0.021 mGy) (p < .0001). Conclusion: CBCT of the wrist carries a significantly higher scattered radiation dose to the neck, chest and abdomen than four view scaphoid radiography, but significantly lower scattered dose than MSCT of the wrist of equivalent diagnostic value.Abstract: Introduction: Scaphoid radiography has poor sensitivity for acute fracture detection and often requires repeat delayed imaging. Although magnetic resonance (MR) imaging is considered the gold standard, computed tomography (CT) is often used as an alternative due to ease of access. Cone-Beam CT (CBCT) offers equivalent diagnostic efficacy to Multi Slice CT (MSCT) at reduced dose. We aimed to establish the difference in scattered dose between modalities for scaphoid imaging. Methods: Anatomical regional entrance surface dose measurements were taken at 3 regions on an anthropomorphic torso phantom positioned as a patient to a wrist phantom undergoing scaphoid imaging for three modalities (CBCT, MSCT, four-view projection radiography). Exposure factors were based on audit of clinical exposures. Each dose measurement was repeated three times per anatomical region, modality, exposure setting and projection. Results: Under unpaired T-test CBCT gave significantly lower mean dose at the neck (1.64 vs 18 mGy), chest (2.78 vs 8.01) and abdomen (1.288 vs 2.93) than MSCT (p < .0001). However CBCT had significantly higher mean dose than four-view radiography at the neck, chest and abdomen (0.031, 0.035, and 0.021 mGy) (p < .0001). Conclusion: CBCT of the wrist carries a significantly higher scattered radiation dose to the neck, chest and abdomen than four view scaphoid radiography, but significantly lower scattered dose than MSCT of the wrist of equivalent diagnostic value. Implications for practice: The use of CBCT for scaphoid injury carries significantly lower scattered dose to radio-sensitive structures investigated here than equivalent MSCT, and may be of greater use as an early cross-sectional investigation for suspected scaphoid fracture. … (more)
- Is Part Of:
- Radiography. Volume 28:Issue 3(2022)
- Journal:
- Radiography
- Issue:
- Volume 28:Issue 3(2022)
- Issue Display:
- Volume 28, Issue 3 (2022)
- Year:
- 2022
- Volume:
- 28
- Issue:
- 3
- Issue Sort Value:
- 2022-0028-0003-0000
- Page Start:
- 811
- Page End:
- 816
- Publication Date:
- 2022-08
- Subjects:
- CBCT -- MSCT -- Scaphoid -- Scatter -- Phantom
Diagnostic imaging -- Periodicals
Radiotherapy -- Periodicals
Cancer -- Radiotherapy -- Periodicals
Diagnostic Imaging -- Periodicals
Neoplasms -- Periodicals
Radiotherapy -- Periodicals
Imagerie pour le diagnostic -- Périodiques
Radiothérapie -- Périodiques
Cancer -- Radiothérapie -- Périodiques
Electronic journals
616.0757 - Journal URLs:
- http://www.sciencedirect.com/science/journal/10788174 ↗
http://www.radiographyonline.com/ ↗
http://www.harcourt-international.com/journals ↗
http://www.idealibrary.com/links/toc/radi/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/10788174 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/10788174 ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiography/ ↗ - DOI:
- 10.1016/j.radi.2022.01.009 ↗
- Languages:
- English
- ISSNs:
- 1078-8174
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7237.001000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 22250.xml