Functionality of registered automated external defibrillators. (July 2022)
- Record Type:
- Journal Article
- Title:
- Functionality of registered automated external defibrillators. (July 2022)
- Main Title:
- Functionality of registered automated external defibrillators
- Authors:
- Jespersen, Sofie Seit
Kjoelbye, Julie Samsoee
Christensen, Helle Collatz
Andelius, Linn
Gregers, Mads Christian Tofte
Torp-Pedersen, Christian
Hansen, Carolina Malta
Folke, Fredrik - Abstract:
- Abstract: Aims: Little is known about automated external defibrillator (AED) functionality in real-life settings. We aimed to assess the functionality of all registered AEDs in a geographically selected area and calculate the proportion of historical out-of-hospital cardiac arrests (OHCAs) covered by non-functioning AEDs. Methods: In this cross-sectional study we inspected all registered and available AEDs on the island of Bornholm in Denmark. We collected information on battery status (determined by AED self-test) and electrode status, as well as AED availability. We identified all historical OHCAs registered with the Danish Cardiac Arrest Registry on Bornholm during 2016–2019 and calculated the proportion of OHCAs covered by an AED (regardless of functionality status) within ≤100, ≤750, and ≤1800 meters and the proportion of OHCAs covered by non-functioning AEDs. Results: Of 211 registered AEDs, 181 (81.9%) were publicly accessible and functional. The remaining 40 (18.1%) were not functional, primarily due to expired electrodes (42.5%, n = 17), obstacles to AED retrieval (20.0%, n = 8) or failed self-tests (17.5%, n = 7). Of 197 historical OHCAs, non-functional AEDs resulted in an OHCA coverage loss of 5.6%, 4.1% and 1.0 % for ≤100 m, ≤750 m and ≤1800 m, respectively. Conclusion: Almost one-fifth of all registered and publicly available AEDs were not functional, primarily due to expired electrodes, failed self-tests or obstacles to retrieving AEDs. One in twentyAbstract: Aims: Little is known about automated external defibrillator (AED) functionality in real-life settings. We aimed to assess the functionality of all registered AEDs in a geographically selected area and calculate the proportion of historical out-of-hospital cardiac arrests (OHCAs) covered by non-functioning AEDs. Methods: In this cross-sectional study we inspected all registered and available AEDs on the island of Bornholm in Denmark. We collected information on battery status (determined by AED self-test) and electrode status, as well as AED availability. We identified all historical OHCAs registered with the Danish Cardiac Arrest Registry on Bornholm during 2016–2019 and calculated the proportion of OHCAs covered by an AED (regardless of functionality status) within ≤100, ≤750, and ≤1800 meters and the proportion of OHCAs covered by non-functioning AEDs. Results: Of 211 registered AEDs, 181 (81.9%) were publicly accessible and functional. The remaining 40 (18.1%) were not functional, primarily due to expired electrodes (42.5%, n = 17), obstacles to AED retrieval (20.0%, n = 8) or failed self-tests (17.5%, n = 7). Of 197 historical OHCAs, non-functional AEDs resulted in an OHCA coverage loss of 5.6%, 4.1% and 1.0 % for ≤100 m, ≤750 m and ≤1800 m, respectively. Conclusion: Almost one-fifth of all registered and publicly available AEDs were not functional, primarily due to expired electrodes, failed self-tests or obstacles to retrieving AEDs. One in twenty historical OHCA was covered by a non-functional AED. Although general AED functionality was high, this finding underlines the importance of regular AED maintenance. … (more)
- Is Part Of:
- Resuscitation. Volume 176(2022)
- Journal:
- Resuscitation
- Issue:
- Volume 176(2022)
- Issue Display:
- Volume 176, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 176
- Issue:
- 2022
- Issue Sort Value:
- 2022-0176-2022-0000
- Page Start:
- 58
- Page End:
- 63
- Publication Date:
- 2022-07
- Subjects:
- Out-of-hospital cardiac arrest -- Automated external defibrillator -- Functionality
Resuscitation -- Periodicals
Resuscitation -- Periodicals
Réanimation -- Périodiques
Electronic journals
616.025 - Journal URLs:
- http://www.sciencedirect.com/science/journal/03009572 ↗
http://www.resuscitationjournal.com/ ↗
http://www.clinicalkey.com/dura/browse/journalIssue/03009572 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/03009572 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.resuscitation.2022.05.013 ↗
- Languages:
- English
- ISSNs:
- 0300-9572
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7785.420000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22105.xml