Three-position characterization for the adjustment of MEMS accelerometer scale factor. (17th December 2020)
- Record Type:
- Journal Article
- Title:
- Three-position characterization for the adjustment of MEMS accelerometer scale factor. (17th December 2020)
- Main Title:
- Three-position characterization for the adjustment of MEMS accelerometer scale factor
- Authors:
- Xu, Tongxu
Xu, Xiang
Bu, Feng
Xu, Dacheng
Zhao, Heming - Abstract:
- Abstract: Up/down static and the six-position least squares calibration methods are commonly used in research conducted on adjustment of the micro-electromechanical systems (MEMS) accelerometer scale factor (or sensitivity). When applying these methods, it is difficult to adjust the accelerometer measuring axis parallel to the gravity direction, causing an error in the accelerometer scale factor. In this paper, a three-position scale factor characterization that does not require corrections of the precise coincident of the measuring axis with the direction of gravity is studied. The method uses the up/down calibration method to obtain the coarse scale factor and bias. The coarse adjustment coefficient is then used to convert the accelerometer output data in the horizontal, up, and down positions into acceleration. The horizontal acceleration is used to estimate the scale factor error. Next, the estimated scale factor error is used to modify the coarse scale factor, producing a more accurate scale factor. The adjustment results from simulation and experiments on triaxial MEMS accelerometers show that the proposed method can effectively correct the scale factor error caused by inability to align the accelerometer measurement axis parallel to the gravity line, with a relative scale factor error of 0.07%, improving the accuracy of the accelerometer scale factor.
- Is Part Of:
- Measurement science & technology. Volume 32:Number 3(2021)
- Journal:
- Measurement science & technology
- Issue:
- Volume 32:Number 3(2021)
- Issue Display:
- Volume 32, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 32
- Issue:
- 3
- Issue Sort Value:
- 2021-0032-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-12-17
- Subjects:
- MEMS accelerometer -- scale factor correction -- scale factor error -- three-position calibration
Physical measurements -- Periodicals
Scientific apparatus and instruments -- Periodicals
Equipment and Supplies -- Periodicals
Science -- instrumentation -- Periodicals
Technology -- instrumentation -- Periodicals
Mesures physiques -- Périodiques
Physical measurements
Scientific apparatus and instruments
Periodicals
502.87 - Journal URLs:
- http://iopscience.iop.org/0957-0233/ ↗
http://www.iop.org/Journals/mt ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1361-6501/abcbce ↗
- Languages:
- English
- ISSNs:
- 0957-0233
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 22068.xml