Effect of MU‐weighted multi‐leaf collimator position error on dose distribution of SBRT radiotherapy in peripheral non‐small cell lung cancer. Issue 12 (31st October 2020)
- Record Type:
- Journal Article
- Title:
- Effect of MU‐weighted multi‐leaf collimator position error on dose distribution of SBRT radiotherapy in peripheral non‐small cell lung cancer. Issue 12 (31st October 2020)
- Main Title:
- Effect of MU‐weighted multi‐leaf collimator position error on dose distribution of SBRT radiotherapy in peripheral non‐small cell lung cancer
- Authors:
- Feng, AiHui
Chen, Hua
Wang, Hao
Gu, HengLe
Shao, Yan
Duan, YanHua
Ying, YanChen
Yue, Ning Jeff
Xu, ZhiYong - Abstract:
- Abstract: Purpose: Position accuracy of the multi‐leaf collimator (MLC) is essential in stereotactic body radiotherapy (SBRT). This study is aimed to investigate the dosimetric impacts of the MU‐weighted MLC positioning uncertainties of SBRT for patients with early stage peripheral non‐small cell lung cancer (NSCLC). Methods: Three types of MLC position error were simulated: Type 1, random error; Type 2, system shift, in which both MLC banks shifted to the left or right direction; and Type 3, in which both MLC banks moved with same magnitudes in the opposite directions. Two baseline plans were generated: an automatic plan (AP) and a manually optimized plan (MP). Multi‐leaf collimator position errors were introduced to generate simulated plans with the preset MLC leaf position errors, which were then reimported into the Pinnacle system to generate simulated plans, respectively. The dosimetric parameters (CI, nCI, GI, etc.) and gEUD values of PTV and OARs were calculated. Linear regression between MU‐weighted/unweighted MLC position error and gEUD was performed to obtain dose sensitivity. Results: The dose sensitivities of the PTVs were −4.93, −38.94, −41.70, −55.55, and 30.33 Gy/mm for random, left shift, right shift, system close, and system open MLC errors, respectively. There were significant differences between the MU‐weighted and the unweighted dose sensitivity, which was −38.94 Gy/mm vs −3.42 Gy/mm (left shift), −41.70 Gy/mm vs −3.56 Gy/mm (right shift), −55.55 Gy/mm vsAbstract: Purpose: Position accuracy of the multi‐leaf collimator (MLC) is essential in stereotactic body radiotherapy (SBRT). This study is aimed to investigate the dosimetric impacts of the MU‐weighted MLC positioning uncertainties of SBRT for patients with early stage peripheral non‐small cell lung cancer (NSCLC). Methods: Three types of MLC position error were simulated: Type 1, random error; Type 2, system shift, in which both MLC banks shifted to the left or right direction; and Type 3, in which both MLC banks moved with same magnitudes in the opposite directions. Two baseline plans were generated: an automatic plan (AP) and a manually optimized plan (MP). Multi‐leaf collimator position errors were introduced to generate simulated plans with the preset MLC leaf position errors, which were then reimported into the Pinnacle system to generate simulated plans, respectively. The dosimetric parameters (CI, nCI, GI, etc.) and gEUD values of PTV and OARs were calculated. Linear regression between MU‐weighted/unweighted MLC position error and gEUD was performed to obtain dose sensitivity. Results: The dose sensitivities of the PTVs were −4.93, −38.94, −41.70, −55.55, and 30.33 Gy/mm for random, left shift, right shift, system close, and system open MLC errors, respectively. There were significant differences between the MU‐weighted and the unweighted dose sensitivity, which was −38.94 Gy/mm vs −3.42 Gy/mm (left shift), −41.70 Gy/mm vs −3.56 Gy/mm (right shift), −55.55 Gy/mm vs −4.84 Gy/mm (system close), and 30.33 vs 2.64 Gy/mm (system open). For the system open/close MLC errors, as the PTV volume became larger, the dose sensitivity decreased. APs provided smaller dose sensitivity for the system shift and system close MLC errors compared to the conventional MPs. Conclusions: There was significant difference in dose sensitivity between MU‐weighted and unweighted MLC position error of SBRT radiotherapy in peripheral NSCLC. MU is suggested to be included in the dosimetric evaluation of the MLC misalignments, since it is much closer to clinical radiotherapy. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 21:Issue 12(2020)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 21:Issue 12(2020)
- Issue Display:
- Volume 21, Issue 12 (2020)
- Year:
- 2020
- Volume:
- 21
- Issue:
- 12
- Issue Sort Value:
- 2020-0021-0012-0000
- Page Start:
- 74
- Page End:
- 83
- Publication Date:
- 2020-10-31
- Subjects:
- auto‐planning -- MLC position error -- MU‐weighted -- SBRT
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13061 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 22030.xml