Development of a scatter correction technique for planar 99mTc-MAA imaging to improve accuracy in lung shunt fraction estimation. (July 2022)
- Record Type:
- Journal Article
- Title:
- Development of a scatter correction technique for planar 99mTc-MAA imaging to improve accuracy in lung shunt fraction estimation. (July 2022)
- Main Title:
- Development of a scatter correction technique for planar 99mTc-MAA imaging to improve accuracy in lung shunt fraction estimation
- Authors:
- McArdle, Niamh
Cournane, Seán
McCavana, Jackie
Lucey, Julie
León Vintró, Luis - Abstract:
- Highlights: Geometric mean method shown to greatly overestimate LSF estimation. Geometric mean method with scatter correction, improves accuracy of LSF estimation. Supported by Monte Carlo, phantom and preliminary patient studies. Revised methodology facilitates improved accuracy in the dosimetry estimation. Facilitates more optimised prescribed patient therapy dose. Abstract: Purpose: Prior to 90 Y selective internal radiation therapy (SIRT) treatment, 99m Tc-MAA scintigraphy imaging is used in the estimation of the lung shunt fraction (LSF). Planar imaging is recommended for determining a LSF ratio. However, the estimate may be affected by scatter contributions, attenuation and respiratory motion. The objective of this study was to correct for the effects of scatter in the LSF, towards the determination of a more accurate estimation method of LSF derived from planar scintigraphy imaging, which is recommended by international guidelines. Methods: The open access SIMIND Monte Carlo modelling software was used to estimate an optimum scatter window (SW) for scatter correction. The uncertainties associated with scatter and scatter contributions from the liver on the LSF were evaluated using an anthropomorphic thorax phantom and a virtual Vox-Man phantom. A brief retrospective examination of patient scans and tumour location investigated the impact that the inclusion of the simulated scatter corrections had on the LSF estimation. Results: The percentage overestimation of theHighlights: Geometric mean method shown to greatly overestimate LSF estimation. Geometric mean method with scatter correction, improves accuracy of LSF estimation. Supported by Monte Carlo, phantom and preliminary patient studies. Revised methodology facilitates improved accuracy in the dosimetry estimation. Facilitates more optimised prescribed patient therapy dose. Abstract: Purpose: Prior to 90 Y selective internal radiation therapy (SIRT) treatment, 99m Tc-MAA scintigraphy imaging is used in the estimation of the lung shunt fraction (LSF). Planar imaging is recommended for determining a LSF ratio. However, the estimate may be affected by scatter contributions, attenuation and respiratory motion. The objective of this study was to correct for the effects of scatter in the LSF, towards the determination of a more accurate estimation method of LSF derived from planar scintigraphy imaging, which is recommended by international guidelines. Methods: The open access SIMIND Monte Carlo modelling software was used to estimate an optimum scatter window (SW) for scatter correction. The uncertainties associated with scatter and scatter contributions from the liver on the LSF were evaluated using an anthropomorphic thorax phantom and a virtual Vox-Man phantom. A brief retrospective examination of patient scans and tumour location investigated the impact that the inclusion of the simulated scatter corrections had on the LSF estimation. Results: The percentage overestimation of the manufacturer recommended method of LSF estimation was 192%. SW corrections improved the uncertainty to within 19% for the range of known LSFs. Similar findings were observed for our patient and tumour location studies. Conclusion: The incorporated scatter corrections can significantly improve the accuracy of the LSF estimation, thereby providing a robust gamma camera, patient and tumour depth specific correction which is easily implementable. This is supported by Monte Carlo, phantom and preliminary patient studies. … (more)
- Is Part Of:
- Physica medica. Volume 99(2022)
- Journal:
- Physica medica
- Issue:
- Volume 99(2022)
- Issue Display:
- Volume 99, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 99
- Issue:
- 2022
- Issue Sort Value:
- 2022-0099-2022-0000
- Page Start:
- 94
- Page End:
- 101
- Publication Date:
- 2022-07
- Subjects:
- Lung shunt fraction -- Scatter correction -- Dosimetry -- Monte Carlo
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2022.05.017 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21963.xml