Non-contact measurement of human respiration using an infrared thermal camera and the deep learning method. (1st July 2022)
- Record Type:
- Journal Article
- Title:
- Non-contact measurement of human respiration using an infrared thermal camera and the deep learning method. (1st July 2022)
- Main Title:
- Non-contact measurement of human respiration using an infrared thermal camera and the deep learning method
- Authors:
- Shu, Shuangbao
Liang, Huajun
Zhang, Yu
Zhang, Yuzhong
Yang, Ziqiang - Abstract:
- Abstract: The existing methods of contact respiration rate (RR) measurement can bring discomfort to the person being measured. However, RR is a human index that has to be monitored in clinical medicine. To overcome the limitations of contact measurement methods, a non-contact RR measurement method based on an infrared thermal camera is proposed. This is based on the phenomenon that human breathing causes periodic temperature changes around the nostrils. First, an infrared thermal camera is used to collect image sequences of the human face. And then, to track a region-of-interest (ROI) in moving image sequences, the You Only Look Once V3 deep learning method is used to track the nostril ROI from the recorded facial image sequences. The performance of the deep learning model is analyzed through experiments. The average temperature of the nostril ROI of the tracked image is calculated, and the temperature change value of the continuous image series can be obtained. The temperature change curve reflects the respiratory process. A Butterworth low-pass filter is used to filter the continuous temperature value to obtain robust respiratory signals, to improve the signal-noise ratio. Last, the respiratory rate is calculated by time-domain and frequency-domain methods. RR is calculated in different conditions, such as normal respiration, fast respiration, and slow respiration. Through the training, the training accuracy of the model reaches 97.9%. Experimental results show that theAbstract: The existing methods of contact respiration rate (RR) measurement can bring discomfort to the person being measured. However, RR is a human index that has to be monitored in clinical medicine. To overcome the limitations of contact measurement methods, a non-contact RR measurement method based on an infrared thermal camera is proposed. This is based on the phenomenon that human breathing causes periodic temperature changes around the nostrils. First, an infrared thermal camera is used to collect image sequences of the human face. And then, to track a region-of-interest (ROI) in moving image sequences, the You Only Look Once V3 deep learning method is used to track the nostril ROI from the recorded facial image sequences. The performance of the deep learning model is analyzed through experiments. The average temperature of the nostril ROI of the tracked image is calculated, and the temperature change value of the continuous image series can be obtained. The temperature change curve reflects the respiratory process. A Butterworth low-pass filter is used to filter the continuous temperature value to obtain robust respiratory signals, to improve the signal-noise ratio. Last, the respiratory rate is calculated by time-domain and frequency-domain methods. RR is calculated in different conditions, such as normal respiration, fast respiration, and slow respiration. Through the training, the training accuracy of the model reaches 97.9%. Experimental results show that the method can achieve high precision, and the maximum error is not more than 2%. Those indicate the proposed method can effectively measure the RR. Therefore, the proposed non-contact RR measurement method can be a useful reference for clinical RR measurement and other applications. … (more)
- Is Part Of:
- Measurement science & technology. Volume 33:Number 7(2022)
- Journal:
- Measurement science & technology
- Issue:
- Volume 33:Number 7(2022)
- Issue Display:
- Volume 33, Issue 7 (2022)
- Year:
- 2022
- Volume:
- 33
- Issue:
- 7
- Issue Sort Value:
- 2022-0033-0007-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-07-01
- Subjects:
- deep learning -- non-contact measurement -- respiration rate measurement -- thermal infrared camera
Physical measurements -- Periodicals
Scientific apparatus and instruments -- Periodicals
Equipment and Supplies -- Periodicals
Science -- instrumentation -- Periodicals
Technology -- instrumentation -- Periodicals
Mesures physiques -- Périodiques
Physical measurements
Scientific apparatus and instruments
Periodicals
502.87 - Journal URLs:
- http://iopscience.iop.org/0957-0233/ ↗
http://www.iop.org/Journals/mt ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1361-6501/ac5ed9 ↗
- Languages:
- English
- ISSNs:
- 0957-0233
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21941.xml