Typical doses and typical values for fluoroscopic diagnostic and interventional procedures. (1st June 2022)
- Record Type:
- Journal Article
- Title:
- Typical doses and typical values for fluoroscopic diagnostic and interventional procedures. (1st June 2022)
- Main Title:
- Typical doses and typical values for fluoroscopic diagnostic and interventional procedures
- Authors:
- Tristram, Juliana
Steuwe, Andrea
Kröpil, Feride
Thomas, Christoph
Rubbert, Christian
Antoch, Gerald
Boos, Johannes - Abstract:
- Abstract: To implement typical doses (TD) and typical values (TV) for fluoroscopic diagnostic and interventional procedures. A total of 3811 fluoroscopic procedures performed within 34 months on three devices were included in this retrospective study. Dose-, patient- and procedure-related information were extracted using the institutional dose management system (DMS). TD/TV were defined as median dose and calculated for the five most frequent procedures per device for dose area product (DAP), cumulative air kerma (CAK) and fluoroscopy time (FT). National diagnostic reference levels and other single facility studies were compared to our results. Additionally, the five procedures with the highest doses of each device were analysed. To evaluate the data coverage of the DMS compared to the picture archiving and communication system (PACS), procedure lists were extracted from the PACS and compared to the procedure information extracted from the DMS. TD/TV for 15 procedures were implemented. Among all devices, TD for DAP ranged between 0.6 Gycm 2 for port catheter control ( n = 64) and 145.9 Gycm 2 for transarterial chemoembolisation ( n = 84). TD for CAK ranged between 5 mGy for port catheter control and 1397 mGy for aneurysm treatment ( n = 129) and TV for FT ranged between 0.3 min for upper cavography ( n = 67) and 51.4 min for aneurysm treatment. TD for DAP and CAK were lower or within the range of other single facility studies. The five procedures with the highest median DAPAbstract: To implement typical doses (TD) and typical values (TV) for fluoroscopic diagnostic and interventional procedures. A total of 3811 fluoroscopic procedures performed within 34 months on three devices were included in this retrospective study. Dose-, patient- and procedure-related information were extracted using the institutional dose management system (DMS). TD/TV were defined as median dose and calculated for the five most frequent procedures per device for dose area product (DAP), cumulative air kerma (CAK) and fluoroscopy time (FT). National diagnostic reference levels and other single facility studies were compared to our results. Additionally, the five procedures with the highest doses of each device were analysed. To evaluate the data coverage of the DMS compared to the picture archiving and communication system (PACS), procedure lists were extracted from the PACS and compared to the procedure information extracted from the DMS. TD/TV for 15 procedures were implemented. Among all devices, TD for DAP ranged between 0.6 Gycm 2 for port catheter control ( n = 64) and 145.9 Gycm 2 for transarterial chemoembolisation ( n = 84). TD for CAK ranged between 5 mGy for port catheter control and 1397 mGy for aneurysm treatment ( n = 129) and TV for FT ranged between 0.3 min for upper cavography ( n = 67) and 51.4 min for aneurysm treatment. TD for DAP and CAK were lower or within the range of other single facility studies. The five procedures with the highest median DAP per device were identified, 6 of 15 procedures were also found to be among the most frequent procedures. Data coverage of the DMS compared to the PACS ranged between 71% (device 2, stroke treatment) and 78% (device 1, lower limb angiography) for the most common procedure per device. Thus, in 22%–29% of cases dose data of the performed procedure was not transferred into the DMS. We implemented TD/TV for fluoroscopic diagnostic and interventional procedures which enable a comprehensive dose analysis and comparison with previously published values. … (more)
- Is Part Of:
- Journal of radiological protection. Volume 42:Number 2(2022)
- Journal:
- Journal of radiological protection
- Issue:
- Volume 42:Number 2(2022)
- Issue Display:
- Volume 42, Issue 2 (2022)
- Year:
- 2022
- Volume:
- 42
- Issue:
- 2
- Issue Sort Value:
- 2022-0042-0002-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-06-01
- Subjects:
- dosimetry -- typical doses -- typical values -- diagnostic reference level -- interventional radiology -- fluoroscopy -- dosimetric comparison
Radiation -- Safety measures -- Periodicals
Health Physics -- Periodicals
Radiation Monitoring -- Periodicals
Radiation Protection -- Periodicals
Rayonnement -- Sécurité -- Mesures -- Périodiques
Electronic journals
Fulltext
Internet Resource
Periodical
363.179905 - Journal URLs:
- http://www.iop.org/EJ/journal/JRP ↗
http://iopscience.iop.org/0952-4746/ ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1361-6498/ac5294 ↗
- Languages:
- English
- ISSNs:
- 0952-4746
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21889.xml