Validation of the blood pressure measurement technology used in the Novacor Diasys 3 (DIS-0001-00) upper arm device for ambulatory blood pressure measurement, according to the requirements of the AAMI/ANSI/ISO 81060-2: 2013 standard (for both a general study and a cardiac-stress study in adults) and of the European Society of Hypertension International Protocol revision 2010. Issue 1 (February 2021)
- Record Type:
- Journal Article
- Title:
- Validation of the blood pressure measurement technology used in the Novacor Diasys 3 (DIS-0001-00) upper arm device for ambulatory blood pressure measurement, according to the requirements of the AAMI/ANSI/ISO 81060-2: 2013 standard (for both a general study and a cardiac-stress study in adults) and of the European Society of Hypertension International Protocol revision 2010. Issue 1 (February 2021)
- Main Title:
- Validation of the blood pressure measurement technology used in the Novacor Diasys 3 (DIS-0001-00) upper arm device for ambulatory blood pressure measurement, according to the requirements of the AAMI/ANSI/ISO 81060-2
- Authors:
- Atkins, Neil
Fania, Claudio
Palatini, Paolo - Abstract:
- Abstract : Objective: The objective of this study was to determine the accuracy of the oscillometric blood pressure measurement technology of the Novacor Diasys 3 (model number DIS-0001-00) blood pressure monitor, intended for ambulatory blood pressure measurement, when used with the recommended Standard (ACC-0213-00), Large (ACC-0214-00) and Paediatric (ACC-0215-00) cuffs. Methods: The Novacor Diasys 3 (model number DIS-0001-00) was evaluated according to the requirements of the AAMI/ANSI/ISO 81060-2:2013 standard, including an additional cardiac-stress study, a requirement for ambulatory blood pressure measurement devices. It was also validated according to the requirements of the European Society of Hypertension International Protocol revision 2010. The protocol requirements for all thee studies were followed precisely. Results: The Novacor Diasys 3 (model number DIS-0001-00) fulfilled all of the requirements for a pass in each of the three studies. In the primary AAMI/ANSI/ISO 81060-2:2013 study, the Criterion 1 errors were +3.9 mmHg ± 2.9 mmHg for SBP and +3.4 ± 2.5 mmHg for DBP. Conclusion: The Novacor Diasys 3 (model number DIS-0001-00), when used with the recommended cuffs, can be recommended for ambulatory blood pressure measurement in the adult population.
- Is Part Of:
- Blood pressure monitoring. Volume 26:Issue 1(2021)
- Journal:
- Blood pressure monitoring
- Issue:
- Volume 26:Issue 1(2021)
- Issue Display:
- Volume 26, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 26
- Issue:
- 1
- Issue Sort Value:
- 2021-0026-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-02
- Subjects:
- accuracy -- ambulatory blood pressure measurement -- cuff -- device -- evaluation -- ESH-IP -- ISO -- validation
Blood pressure -- Measurement -- Periodicals
Blood Pressure Determination -- Periodicals
Blood Pressure Monitoring, Ambulatory -- Periodicals
Blood Pressure Monitors -- Periodicals
Hypertension -- Periodicals
612.140287 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&PAGE=toc&D=ovft&AN=00126097-000000000-00000&NEWS=N ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00126097-000000000-00000 ↗
http://www.bpmonitoring.com/ ↗
http://journals.lww.com/pages/default.aspx ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1097/MBP.0000000000000493 ↗
- Languages:
- English
- ISSNs:
- 1359-5237
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2113.035500
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21891.xml