Fiducial marker position affects target volume in stereotactic lung irradiation. Issue 6 (4th April 2022)
- Record Type:
- Journal Article
- Title:
- Fiducial marker position affects target volume in stereotactic lung irradiation. Issue 6 (4th April 2022)
- Main Title:
- Fiducial marker position affects target volume in stereotactic lung irradiation
- Authors:
- Akasaka, Hiroaki
Mizonobe, Kazufusa
Oki, Yuya
Uehara, Kazuyuki
Nakayama, Masao
Tamura, Shuhei
Munetomo, Yoshiki
Kawaguchi, Haruna
Ishida, Jun
Harada, Aya
Ishihara, Takeaki
Kubota, Hikaru
Kawaguchi, Hiroki
Sasaki, Ryohei
Mayahara, Hiroshi - Abstract:
- Abstract: Purpose: Real‐time tracking systems of moving respiratory targets such as CyberKnife, Radixact, or Vero4DRT are an advanced robotic radiotherapy device used to deliver stereotactic body radiotherapy (SBRT). The internal target volume (ITV) of lung tumors is assessed through a fiducial marker fusion using four‐dimensional computed tomography (CT). It is important to minimize the ITV to protect normal lung tissue from exposure to radiation and the associated side effects post SBRT. However, the ITV may alter if there is a change in the position of the fiducial marker with respect to the tumor. This study investigated the relationship between fiducial marker position and the ITV in order to prevent radiation exposure of normal lung tissue, and correct target coverage. Materials and methods: This study retrospectively reviewed 230 lung cancer patients who received a fiducial marker for SBRT between April 2015 and September 2021. The distance of the fiducial marker to the gross tumor volume (GTV) in the expiratory ( d ex ) and inspiratory ( d in ) CT, and the ratio of the ITV/V(GTVex ), were investigated. Results: Upon comparing each lobe, although there was no significant difference in the d diff and the ITV/V(GTVex ) between all lobes for d ex < 10 mm, there was significant difference in the d diff and the ITV/V(GTVex ) between the lower and upper lobes for d ex ≥ 10 mm ( p < 0.05). Moreover, there was significant difference in the d diff and the ITV/V(GTVex )Abstract: Purpose: Real‐time tracking systems of moving respiratory targets such as CyberKnife, Radixact, or Vero4DRT are an advanced robotic radiotherapy device used to deliver stereotactic body radiotherapy (SBRT). The internal target volume (ITV) of lung tumors is assessed through a fiducial marker fusion using four‐dimensional computed tomography (CT). It is important to minimize the ITV to protect normal lung tissue from exposure to radiation and the associated side effects post SBRT. However, the ITV may alter if there is a change in the position of the fiducial marker with respect to the tumor. This study investigated the relationship between fiducial marker position and the ITV in order to prevent radiation exposure of normal lung tissue, and correct target coverage. Materials and methods: This study retrospectively reviewed 230 lung cancer patients who received a fiducial marker for SBRT between April 2015 and September 2021. The distance of the fiducial marker to the gross tumor volume (GTV) in the expiratory ( d ex ) and inspiratory ( d in ) CT, and the ratio of the ITV/V(GTVex ), were investigated. Results: Upon comparing each lobe, although there was no significant difference in the d diff and the ITV/V(GTVex ) between all lobes for d ex < 10 mm, there was significant difference in the d diff and the ITV/V(GTVex ) between the lower and upper lobes for d ex ≥ 10 mm ( p < 0.05). Moreover, there was significant difference in the d diff and the ITV/V(GTVex ) between d ex ≥10 mm and d ex < 10 mm in all lung regions ( p < 0.05). Conclusion: The ITV that had no margin from GTVs increased when d ex was ≥10 mm for all lung regions ( p < 0.05). Furthermore, the increase in ITV tended to be greater in the lower lung lobe. These findings can help decrease the possibility of adverse events post SBRT, and correct target coverage. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 23:Issue 6(2022)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 23:Issue 6(2022)
- Issue Display:
- Volume 23, Issue 6 (2022)
- Year:
- 2022
- Volume:
- 23
- Issue:
- 6
- Issue Sort Value:
- 2022-0023-0006-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2022-04-04
- Subjects:
- CyberKnife -- fiducial marker -- internal target volume -- lung cancer -- radiotherapy -- stereotactic body radiotherapy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13596 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21812.xml