Chest computer tomography is safe without additional interrogation or monitoring for modern cardiac implantable electrical devices. (8th April 2022)
- Record Type:
- Journal Article
- Title:
- Chest computer tomography is safe without additional interrogation or monitoring for modern cardiac implantable electrical devices. (8th April 2022)
- Main Title:
- Chest computer tomography is safe without additional interrogation or monitoring for modern cardiac implantable electrical devices
- Authors:
- Tsutsui, Kenta
Kato, Ritsushi
Mori, Hitoshi
Kawano, Daisuke
Ikeda, Yoshifumi
Matsumoto, Kei
Tanaka, Naomichi
Hoya, Hiroki
Seki, Kaname
Yokoyama, Shinji
Yoshida, Yuzuru
Iida, Shinichiro
Yamamoto, Keiji
Nakano, Shintaro
Iwanaga, Shiro
Muramatsu, Toshihiro
Matsumoto, Kazuo - Abstract:
- Abstract: Introduction: Whether diagnostic computed tomography (CT) scans to cardiac implantable electronic devices (CIED) is safe in recent models remains unknown. Methods: A two‐centers observational study. Over 14 years, consecutive 2362 chest CT scans (1666 pacemakers [PMs], 145 cardiac resynchronization therapy PM, 316 implantable cardioverter‐defibrillator, and 233 cardiac resynchronization therapy defibrillator) were interrogated and monitored upon imaging. Results: Electromagnetic interference occurred only in a few old models: InSync 8040 ( n = 14), InSync III Marquis ( n = 1), and Kappa ( n = 4), which resulted no adverse events. Conclusion: CIEDs, especially recent ones, are confirmed safe on chest CT.
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 33:Number 6(2022)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 33:Number 6(2022)
- Issue Display:
- Volume 33, Issue 6 (2022)
- Year:
- 2022
- Volume:
- 33
- Issue:
- 6
- Issue Sort Value:
- 2022-0033-0006-0000
- Page Start:
- 1336
- Page End:
- 1340
- Publication Date:
- 2022-04-08
- Subjects:
- CIEDs -- chest CT scan -- electromagnetic interference
Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.15475 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21816.xml