Modified snare technique improves left ventricular lead implant success for cardiac resynchronization therapy. (28th September 2020)
- Record Type:
- Journal Article
- Title:
- Modified snare technique improves left ventricular lead implant success for cardiac resynchronization therapy. (28th September 2020)
- Main Title:
- Modified snare technique improves left ventricular lead implant success for cardiac resynchronization therapy
- Authors:
- Marques, Pedro
Nunes‐Ferreira, Afonso
António, Pedro S.
Aguiar‐Ricardo, Inês
Lima da Silva, Gustavo
Guimarães, Tatiana
Bernardes, Ana
Santos, Igor
Pinto, Fausto J.
de Sousa, João - Abstract:
- Abstract: Background: Left ventricular (LV) lead placement is the most challenging aspect of cardiac resynchronization therapy (CRT) device implantation, with a failure rate of up to 10% due to complex coronary anatomies. We describe a modified snare technique for LV lead placement and evaluate its safety and efficacy in cases when standard methods fail. Methods and Results: A prospective study was conducted of patients indicated for a CRT implant. When LV lead delivery to the target vessel failed using standard techniques, a modified snare technique was employed. Patients were evaluated every 6 months. From 2015 to 2019, 566 CRTs were implanted (26.1% female, 72 ± 10.2 years old, follow‐up duration 18.9 ± 15.8 months). The standard LV implant technique failed in 94 cases (16.6%), of which the modified snare technique was successful in 92 (97.9%). There were no differences between the modified snare and standard techniques in the rates of 30‐day postimplant CRT all‐cause mortality (3.2% vs. 1.7%, p = .33), 4‐year all‐cause mortality (15.9% vs. 15.5%, p = .49), or major acute complications (7.4% vs. 3.8%, p = .12). However, the 4‐year procedural reintervention rate was lower with the modified snare technique (3.2% vs. 10.2%, p < .05), specifically LV implant failure or dislodgement rates (0% vs. 5.3%, p < .05), improving the response rate (71.8% vs. 55.1%, p < .05). Conclusions: For challenging coronary sinus anatomies that preclude LV lead placement by standardAbstract: Background: Left ventricular (LV) lead placement is the most challenging aspect of cardiac resynchronization therapy (CRT) device implantation, with a failure rate of up to 10% due to complex coronary anatomies. We describe a modified snare technique for LV lead placement and evaluate its safety and efficacy in cases when standard methods fail. Methods and Results: A prospective study was conducted of patients indicated for a CRT implant. When LV lead delivery to the target vessel failed using standard techniques, a modified snare technique was employed. Patients were evaluated every 6 months. From 2015 to 2019, 566 CRTs were implanted (26.1% female, 72 ± 10.2 years old, follow‐up duration 18.9 ± 15.8 months). The standard LV implant technique failed in 94 cases (16.6%), of which the modified snare technique was successful in 92 (97.9%). There were no differences between the modified snare and standard techniques in the rates of 30‐day postimplant CRT all‐cause mortality (3.2% vs. 1.7%, p = .33), 4‐year all‐cause mortality (15.9% vs. 15.5%, p = .49), or major acute complications (7.4% vs. 3.8%, p = .12). However, the 4‐year procedural reintervention rate was lower with the modified snare technique (3.2% vs. 10.2%, p < .05), specifically LV implant failure or dislodgement rates (0% vs. 5.3%, p < .05), improving the response rate (71.8% vs. 55.1%, p < .05). Conclusions: For challenging coronary sinus anatomies that preclude LV lead placement by standard methods, this modified snare alternative was safe and effective, with comparable mortality and complications, but significantly lower procedural reintervention and higher response rates. … (more)
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 31:Number 11(2020)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 31:Number 11(2020)
- Issue Display:
- Volume 31, Issue 11 (2020)
- Year:
- 2020
- Volume:
- 31
- Issue:
- 11
- Issue Sort Value:
- 2020-0031-0011-0000
- Page Start:
- 2954
- Page End:
- 2963
- Publication Date:
- 2020-09-28
- Subjects:
- cardiac resynchronization therapy -- efficacy -- left ventricular lead -- responders -- safety -- snare technique
Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.14750 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21705.xml