Proton dose calculation based on converting dual-energy CT data to stopping power ratio (DEEDZ-SPR): a beam-hardening assessment. (16th December 2020)
- Record Type:
- Journal Article
- Title:
- Proton dose calculation based on converting dual-energy CT data to stopping power ratio (DEEDZ-SPR): a beam-hardening assessment. (16th December 2020)
- Main Title:
- Proton dose calculation based on converting dual-energy CT data to stopping power ratio (DEEDZ-SPR): a beam-hardening assessment
- Authors:
- Tanaka, Sodai
Noto, Yoshiyuki
Utsunomiya, Satoru
Yoshimura, Takaaki
Matsuura, Taeko
Saito, Masatoshi - Abstract:
- Abstract: To achieve an accurate stopping power ratio (SPR) prediction in particle therapy treatment planning, we previously proposed a simple conversion to the SPR from dual-energy (DE) computed tomography (CT) data via electron density and effective atomic number ( Z eff ) calibration (DEEDZ-SPR). This study was conducted to carry out an initial implementation of the DEEDZ-SPR conversion method with a clinical treatment planning system (TPS; VQA, Hitachi Ltd., Tokyo) for proton beam therapy. Consequently, this paper presents a proton therapy plan for an anthropomorphic phantom to evaluate the stability of the dose calculations obtained by the DEEDZ-SPR conversion against the variation of the calibration phantom size. Dual-energy x-ray CT images were acquired using a dual-source CT (DSCT) scanner. A single-energy CT (SECT) scan using the same DSCT scanner was also performed to compare the DEEDZ-SPR conversion with the SECT-based SPR (SECT-SPR) conversion. The scanner-specific parameters necessary for the SPR calibration were obtained from the CT images of tissue substitutes in a calibration phantom. Two calibration phantoms with different sizes (a 33 cm diameter phantom and an 18 cm diameter phantom) were used for the SPR calibrations to investigate the beam-hardening effect on dosimetric uncertainties. Each set of calibrated SPR data was applied to the proton therapy plan designed using the VQA TPS with a pencil beam algorithm for the anthropomorphic phantom. The treatmentAbstract: To achieve an accurate stopping power ratio (SPR) prediction in particle therapy treatment planning, we previously proposed a simple conversion to the SPR from dual-energy (DE) computed tomography (CT) data via electron density and effective atomic number ( Z eff ) calibration (DEEDZ-SPR). This study was conducted to carry out an initial implementation of the DEEDZ-SPR conversion method with a clinical treatment planning system (TPS; VQA, Hitachi Ltd., Tokyo) for proton beam therapy. Consequently, this paper presents a proton therapy plan for an anthropomorphic phantom to evaluate the stability of the dose calculations obtained by the DEEDZ-SPR conversion against the variation of the calibration phantom size. Dual-energy x-ray CT images were acquired using a dual-source CT (DSCT) scanner. A single-energy CT (SECT) scan using the same DSCT scanner was also performed to compare the DEEDZ-SPR conversion with the SECT-based SPR (SECT-SPR) conversion. The scanner-specific parameters necessary for the SPR calibration were obtained from the CT images of tissue substitutes in a calibration phantom. Two calibration phantoms with different sizes (a 33 cm diameter phantom and an 18 cm diameter phantom) were used for the SPR calibrations to investigate the beam-hardening effect on dosimetric uncertainties. Each set of calibrated SPR data was applied to the proton therapy plan designed using the VQA TPS with a pencil beam algorithm for the anthropomorphic phantom. The treatment plans with the SECT-SPR conversion exhibited discrepancies between the dose distributions and the dose-volume histograms (DVHs) of the 33 cm and 18 cm phantom calibrations. In contrast, the corresponding dose distributions and the DVHs obtained using the DEEDZ-SPR conversion method coincided almost perfectly with each other. The DEEDZ-SPR conversion appears to be a promising method for providing proton dose plans that are stable against the size variations of the calibration phantom and the patient. … (more)
- Is Part Of:
- Physics in medicine & biology. Volume 65:Number 23(2020:Dec.)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 65:Number 23(2020:Dec.)
- Issue Display:
- Volume 65, Issue 23 (2020)
- Year:
- 2020
- Volume:
- 65
- Issue:
- 23
- Issue Sort Value:
- 2020-0065-0023-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-12-16
- Subjects:
- stopping power ratio -- dual-energy CT imaging -- proton therapy -- treatment planning -- dose distribution
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/abae09 ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21690.xml