Efficient 23Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum 23Na (CRISTINA). Issue 5 (28th May 2020)
- Record Type:
- Journal Article
- Title:
- Efficient 23Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum 23Na (CRISTINA). Issue 5 (28th May 2020)
- Main Title:
- Efficient 23Na triple‐quantum signal imaging on clinical scanners: Cartesian imaging of single and triple‐quantum 23Na (CRISTINA)
- Authors:
- Hoesl, Michaela A. U.
Schad, Lothar R.
Rapacchi, Stanislas - Abstract:
- Abstract : Purpose: To capture the multiquantum coherence (MQC) 23 Na signal. Different phase‐cycling options and sequences are compared in a unified theoretical layout, and a novel sequence is developed. Methods: An open source simulation overview is provided with graphical explanations to facilitate MQC understanding and access to techniques. Biases such as B0 inhomogeneity and stimulated echo signal were simulated for 4 different phase‐cycling options previously described. Considerations for efficiency and accuracy lead to the implementation of a 2D Cartesian single and triple quantum imaging of sodium (CRISTINA) sequence employing two 6‐step cycles in combination with a multi‐echo readout. CRISTINA was compared to simultaneous single‐quantum and triple‐quantum‐filtered MRI of sodium (SISTINA) under strong static magnetic gradient. CRISTINA capabilities were assessed on 8 × 60 mL, 0% to 5% agarose phantom with 50 to 154 mM 23 Na concentration at 7 T. CRISTINA was demonstrated subsequently in vivo in the brain. Results: Simulation of B0 inhomogeneity showed severe signal dropout, which can lead to erroneous MQC measurement. Stimulated echo signal was highest at the time of triple‐quantum coherences signal maximum. However, stimulated echo signal is separated by Fourier Transform as an offset and did not interfere with MQC signals. The multi‐echo readout enabled capturing both single‐quantum coherences and triple‐quantum coherences signal evolution at once. SignalAbstract : Purpose: To capture the multiquantum coherence (MQC) 23 Na signal. Different phase‐cycling options and sequences are compared in a unified theoretical layout, and a novel sequence is developed. Methods: An open source simulation overview is provided with graphical explanations to facilitate MQC understanding and access to techniques. Biases such as B0 inhomogeneity and stimulated echo signal were simulated for 4 different phase‐cycling options previously described. Considerations for efficiency and accuracy lead to the implementation of a 2D Cartesian single and triple quantum imaging of sodium (CRISTINA) sequence employing two 6‐step cycles in combination with a multi‐echo readout. CRISTINA was compared to simultaneous single‐quantum and triple‐quantum‐filtered MRI of sodium (SISTINA) under strong static magnetic gradient. CRISTINA capabilities were assessed on 8 × 60 mL, 0% to 5% agarose phantom with 50 to 154 mM 23 Na concentration at 7 T. CRISTINA was demonstrated subsequently in vivo in the brain. Results: Simulation of B0 inhomogeneity showed severe signal dropout, which can lead to erroneous MQC measurement. Stimulated echo signal was highest at the time of triple‐quantum coherences signal maximum. However, stimulated echo signal is separated by Fourier Transform as an offset and did not interfere with MQC signals. The multi‐echo readout enabled capturing both single‐quantum coherences and triple‐quantum coherences signal evolution at once. Signal combination of 2 phase‐cycles with a corresponding B0 map was found to recover the signal optimally. Experimental results confirm and complement the simulations. Conclusion: Considerations for efficient MQC measurements, most importantly avoiding B0 signal loss, led to the design of CRISTINA. CRISTINA captures triple‐quantum coherences and single‐quantum coherences signal evolution to provide complete sodium signal characterization including T 2 ∗ fast, T 2 ∗ slow, MQC amplitudes, and sodium concentration. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 84:Issue 5(2020)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 84:Issue 5(2020)
- Issue Display:
- Volume 84, Issue 5 (2020)
- Year:
- 2020
- Volume:
- 84
- Issue:
- 5
- Issue Sort Value:
- 2020-0084-0005-0000
- Page Start:
- 2412
- Page End:
- 2428
- Publication Date:
- 2020-05-28
- Subjects:
- 23Na MRI -- coherence transfer -- multiple‐quantum coherences -- triple‐quantum sodium imaging
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.28284 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21625.xml