A FE study on the effect of interference fit and coefficient of friction on the micromotions and interface gaps of a cementless PEEK femoral component. (May 2022)
- Record Type:
- Journal Article
- Title:
- A FE study on the effect of interference fit and coefficient of friction on the micromotions and interface gaps of a cementless PEEK femoral component. (May 2022)
- Main Title:
- A FE study on the effect of interference fit and coefficient of friction on the micromotions and interface gaps of a cementless PEEK femoral component
- Authors:
- Post, Corine E.
Bitter, Thom
Briscoe, Adam
Verdonschot, Nico
Janssen, Dennis - Abstract:
- Abstract: The use of a more compliant material, such as polyetheretherketone (PEEK), for a cementless femoral component is a potential solution to prevent aseptic loosening caused by peri-prosthetic stress-shielding. Long-term fixation of a cementless femoral component is achieved by a proper primary fixation of the bone-implant interface, which is influenced by the interference fit and frictional properties of the implant surface. This computational study investigates the sensitivity of micromotions and interface gaps of a cementless PEEK femoral component to the interference fit and coefficient of friction. 24 finite element models of the femur and femoral component were created with variations in implant material, interference fit and coefficient of friction. Peak loads of a jogging activity were applied on the models. Micromotions and interface gaps were both sensitive to the interference fit, coefficient of friction and implant material. Besides the implant material, the micromotions and interface gaps of the implant were most sensitive to the interference fit. Compared to the cobalt-chrome (CoCr) femoral component, the PEEK femoral component generated higher micromotions and interface gaps when equal interference fit and friction values were applied. However, increasing the interference fit and friction of the PEEK component resulted in micromotion values comparable with the CoCr component. This result leads to possibilities using cementless PEEK femoral components.
- Is Part Of:
- Journal of biomechanics. Volume 137(2022)
- Journal:
- Journal of biomechanics
- Issue:
- Volume 137(2022)
- Issue Display:
- Volume 137, Issue 2022 (2022)
- Year:
- 2022
- Volume:
- 137
- Issue:
- 2022
- Issue Sort Value:
- 2022-0137-2022-0000
- Page Start:
- Page End:
- Publication Date:
- 2022-05
- Subjects:
- Total knee arthroplasty -- Polyetheretherketone -- Uncemented femoral knee component -- Finite element simulation -- Micromotions
Animal mechanics -- Periodicals
Biomechanics -- Periodicals
Biomechanics -- Periodicals
Mécanique animale -- Périodiques
Biomécanique -- Périodiques
Electronic journals
571.4305 - Journal URLs:
- http://www.sciencedirect.com/science/journal/00219290 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/00219290 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/00219290 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jbiomech.2022.111057 ↗
- Languages:
- English
- ISSNs:
- 0021-9290
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.600000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21528.xml