Laboratory methods for monitoring argatroban in heparin‐induced thrombocytopenia. (9th November 2021)
- Record Type:
- Journal Article
- Title:
- Laboratory methods for monitoring argatroban in heparin‐induced thrombocytopenia. (9th November 2021)
- Main Title:
- Laboratory methods for monitoring argatroban in heparin‐induced thrombocytopenia
- Authors:
- Guy, Susan
Kitchen, Steve
Hopkins, Barbara
Chunara, Zunaid
Stephenson‐Brown, Alex
van Veen, Joost J. - Abstract:
- Abstract: Introduction: The Summary of Product Characteristics for the direct thrombin inhibitor argatroban states monitoring should be by activated partial thromboplastin time (APTT), with a target range of 1.5‐3.0 times the patients' baseline APTT. APTT may be influenced by coagulopathies, lupus anticoagulant and raised FVIII levels. Previous studies have shown sensitivity differences of APTT reagents to argatroban. Some recent publications have favoured the use of anti‐IIa methods to determine the plasma drug concentration of argatroban. This study aims to compare the anti‐IIa assays: Hemoclot thrombin inhibitor assay (HTI) and Ecarin chromogenic assay (ECA) alongside the APTT. Methods: Residual plasma of 25 samples from 8 patients (3 with COVID‐19 and HIT: n = 18, 5 with HIT: n = 7) was tested at two sites: site A: Sysmex CS5100 by HTI and APTT (Actin FS and SynthASil), and also on Stago STA Compact Max: ECA and APTT (CK Prest); and site B: Stago STA R Max 2 by ECA and APTT (Cephascreen). Results: Mean APTT ratio was 1.96 (Actin FS), 1.84 (SynthASil), 1.59 (CK Prest) and 2.48 (Cephascreen). Mean argatroban concentration by HTI was 0.60 µg/mL and by ECA was 0.65 µg/mL (site A) and 0.70 µg/mL (site B). There was a poor correlation to HTI in APTT ratios (range r 2 = .0235‐0.4181) with stronger correlations between ECA methods to HTI ( r 2 = .8998 site A, r 2 = .8734 site B). Conclusion: This study confirms previous publications on the unsuitability of APTT and expandsAbstract: Introduction: The Summary of Product Characteristics for the direct thrombin inhibitor argatroban states monitoring should be by activated partial thromboplastin time (APTT), with a target range of 1.5‐3.0 times the patients' baseline APTT. APTT may be influenced by coagulopathies, lupus anticoagulant and raised FVIII levels. Previous studies have shown sensitivity differences of APTT reagents to argatroban. Some recent publications have favoured the use of anti‐IIa methods to determine the plasma drug concentration of argatroban. This study aims to compare the anti‐IIa assays: Hemoclot thrombin inhibitor assay (HTI) and Ecarin chromogenic assay (ECA) alongside the APTT. Methods: Residual plasma of 25 samples from 8 patients (3 with COVID‐19 and HIT: n = 18, 5 with HIT: n = 7) was tested at two sites: site A: Sysmex CS5100 by HTI and APTT (Actin FS and SynthASil), and also on Stago STA Compact Max: ECA and APTT (CK Prest); and site B: Stago STA R Max 2 by ECA and APTT (Cephascreen). Results: Mean APTT ratio was 1.96 (Actin FS), 1.84 (SynthASil), 1.59 (CK Prest) and 2.48 (Cephascreen). Mean argatroban concentration by HTI was 0.60 µg/mL and by ECA was 0.65 µg/mL (site A) and 0.70 µg/mL (site B). There was a poor correlation to HTI in APTT ratios (range r 2 = .0235‐0.4181) with stronger correlations between ECA methods to HTI ( r 2 = .8998 site A, r 2 = .8734 site B). Conclusion: This study confirms previous publications on the unsuitability of APTT and expands this by being multisited and included APTT reagents that use mechanical clot detection. Both anti‐IIa methods are more suitable. … (more)
- Is Part Of:
- International journal of laboratory hematology. Volume 44:Number 2(2022)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 44:Number 2(2022)
- Issue Display:
- Volume 44, Issue 2 (2022)
- Year:
- 2022
- Volume:
- 44
- Issue:
- 2
- Issue Sort Value:
- 2022-0044-0002-0000
- Page Start:
- 399
- Page End:
- 406
- Publication Date:
- 2021-11-09
- Subjects:
- APTT -- argatroban -- dilute thrombin time -- Ecarin chromogenic assay -- HIT
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.13757 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 21525.xml