Evaluation of Hepatic Iron Overload Using a Contemporary 0.55 T MRI System. Issue 6 (20th October 2021)
- Record Type:
- Journal Article
- Title:
- Evaluation of Hepatic Iron Overload Using a Contemporary 0.55 T MRI System. Issue 6 (20th October 2021)
- Main Title:
- Evaluation of Hepatic Iron Overload Using a Contemporary 0.55 T MRI System
- Authors:
- Campbell‐Washburn, Adrienne E.
Mancini, Christine
Conrey, Anna
Edwards, Lanelle
Shanbhag, Sujata
Wood, John
Xue, Hui
Kellman, Peter
Bandettini, W. Patricia
Thein, Swee Lay - Abstract:
- Abstract : Background: MRI T2* and R2* mapping have gained clinical acceptance for noninvasive assessment of iron overload. Lower field MRI may offer increased measurement dynamic range in patients with high iron concentration and may potentially increase MRI accessibility, but it is compromised by lower signal‐to‐noise ratio that reduces measurement precision. Purpose: To characterize a high‐performance 0.55 T MRI system for evaluating patients with liver iron overload. Study Type: Prospective. Population: Forty patients with known or suspected iron overload (sickle cell anemia [ n = 5], ß‐ thalassemia [ n = 3], and hereditary spherocytosis [ n = 2]) and a liver iron phantom. Field strength/sequence: A breath‐held multiecho gradient echo sequence at 0.55 T and 1.5 T. Assessment: Patients were imaged with T2*/R2* mapping 0.55 T and 1.5 T within 24 hours, and 16 patients returned for follow‐up exams within 6–16 months, resulting in 56 paired studies. Liver T2* and R2* measurements and standard deviations were compared between 0.55 T and 1.5 T and used to validate a predictive model between field strengths. The model was then used to classify iron overload at 0.55 T. Statistical Tests: Linear regression and Bland–Altman analysis were used for comparisons, and measurement precision was assessed using the coefficient of variation. A P ‐value < 0.05 was considered statistically significant. Results: R2* was significantly lower at 0.55 T in our cohort (488 ± 449 s −1 at 1.5 TAbstract : Background: MRI T2* and R2* mapping have gained clinical acceptance for noninvasive assessment of iron overload. Lower field MRI may offer increased measurement dynamic range in patients with high iron concentration and may potentially increase MRI accessibility, but it is compromised by lower signal‐to‐noise ratio that reduces measurement precision. Purpose: To characterize a high‐performance 0.55 T MRI system for evaluating patients with liver iron overload. Study Type: Prospective. Population: Forty patients with known or suspected iron overload (sickle cell anemia [ n = 5], ß‐ thalassemia [ n = 3], and hereditary spherocytosis [ n = 2]) and a liver iron phantom. Field strength/sequence: A breath‐held multiecho gradient echo sequence at 0.55 T and 1.5 T. Assessment: Patients were imaged with T2*/R2* mapping 0.55 T and 1.5 T within 24 hours, and 16 patients returned for follow‐up exams within 6–16 months, resulting in 56 paired studies. Liver T2* and R2* measurements and standard deviations were compared between 0.55 T and 1.5 T and used to validate a predictive model between field strengths. The model was then used to classify iron overload at 0.55 T. Statistical Tests: Linear regression and Bland–Altman analysis were used for comparisons, and measurement precision was assessed using the coefficient of variation. A P ‐value < 0.05 was considered statistically significant. Results: R2* was significantly lower at 0.55 T in our cohort (488 ± 449 s −1 at 1.5 T vs. 178 ± 155 s −1 at 0.55 T, n = 56 studies) and in the patients with severe iron overload (937 ± 369 s −1 at 1.5 T vs. 339 ± 127 s −1 at 0.55 T, n = 23 studies). The coefficient of variation indicated reduced precision at 0.55 T (3.5 ± 2.2% at 1.5 T vs 6.9 ± 3.9% at 0.55 T). The predictive model accurately predicted 1.5 T R2* from 0.55 T R2* (Bland Altman bias = −6.6 ± 20.5%). Using this model, iron overload at 0.55 T was classified as: severe R2* > 185 s −1, moderate 81 s −1 < R2* < 185 s −1, and mild 45 s −1 < R2* < 91 s −1 . Data Conclusion: We demonstrated that 0.55 T provides T2* and R2* maps that can be used for the assessment of liver iron overload in patients. Evidence Level: 2 Technical Efficacy: Stage 2 … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 55:Issue 6(2022)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 55:Issue 6(2022)
- Issue Display:
- Volume 55, Issue 6 (2022)
- Year:
- 2022
- Volume:
- 55
- Issue:
- 6
- Issue Sort Value:
- 2022-0055-0006-0000
- Page Start:
- 1855
- Page End:
- 1863
- Publication Date:
- 2021-10-20
- Subjects:
- iron overload -- T2* -- R2* -- low field -- MRI
Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.27950 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 21496.xml